9
No.
Instruction
1
First-level operation directory, through the “↑↓” key to select these functions,
when these functions are selected, press the “←”→”" key to select the
parameters of the corresponding functions.
Tip: The cursor can quickly jump to the button by pressing the "Enter"
button under the first-level operation directory at “Start Test”.
No. of
tested
product
Set the number of tested product.
Test
winding
Set the test winding of the tested product.
No.
Instruction
Tapping
position
Sets the current tap position.
Test
phase
Set the phase of the tested product
,
can select“single phase AO”
、
“single
phase BO”
、
“single phase CO”
、
“single phase AB”
、
“single phase BC”
、
“single
phase CA”
、
“three phase Yn”
、
“three phase Y/
△
”
;
Note:
Three-phase unbalance rate can be calculated automatically when
three-phase testing.
Winding
materials
Set the winding material of the tested product, copper and aluminum are
optional. The winding material relates to the conversion coefficient used for
the resistance conversion value.
Test
temperat
ure
Set the current temperature of the tested product, and the temperature
value is from - 99
℃
to + 99
℃
. The device can monitor the oil temperature
inside the test product in real time through an external wireless temperature
measuring probe, or set the temperature value manually. When the external
wireless temperature probe is connected, it can't be set manually, so it
needs to close the probe to set manually.
Convert
temperat
ure
Setting the measured resistance value needs to be converted to the
temperature value. The converted temperature value is from 0
℃
to + 255
℃
,
which is related to the accuracy of the converted resistance value.
No.
Instruction
Testing
current
Select the test current gear
,
Choose
1mA
、
10mA
、
0.1A
、
1A
、
5A
、
10A
and
automatic.
2
The secondary operation directory corresponds to the setting parameters of
the primary operation directory, and the parameters are modified by “←”→”
key to move the cursor and “↑↓” key.
Tip: the cursor in the second level of operation directory, you can press
the "Enter" or "Cancel" key to jump the cursor to the first level of operation
directory.
3
When the cursor is here, press “Enter” to start the measurement.
4
The range of current selected for testing.
5
An explanation of the selected function.