8

Select the "DC Resistance Testing" menu to enter the DC resistance parameter

settings screen, as shown below.

No.

Instruction

1

Display date and time.

2

Displays the peripheral and current operation status.
This icon is displayed when you insert a USB drive.
This icon is displayed when the printer is inserted.
This icon is displayed when the temperature measurement module is
connected.
This icon is displayed during testing.
Temperature overheating inside the meter display this icon

3

Power display. This icon flashes when the power is low.

No.

Instruction

4

The main menu operation area of the instrument, select corresponding
functions through “←”→” keys, and press "Enter" to enter the
corresponding menu.
DC resistance test function, can measure transformer, mutual inductor,
reactor and other inductive test products and copper bars, conductors,
switch contacts and other resistance test products.
In the process of query and test, each group of data can be saved; in
the storage query screen, data can be printed and transferred to the
USB disk.
Set the date and time of the instrument.
Password operation is required, not available to the user.

5

SV

Displays the current software version number of the instrument

HV

Displays the current hardware version number of the instrument.

Summary of Contents for HZ-3310E-I

Page 1: ...HZ 3310E I Portable DC Winding Resistance Tester Huazheng Electric Manufacturing Baoding Co Ltd Dear user...

Page 2: ...l and operate and maintain the instrument according to the manual to prolong its service life Just a light press the test will be completed automatically is the operating characteristics of this instr...

Page 3: ...aging Content 1 IV Functional Characteristics 2 V Technical Indicators 3 VI Adjust the Wristband 3 VII Battery Charging 4 VIII Tilt Hand held Tester 4 IX Product Appearance 4 X Operation Instructions...

Page 4: ...return to the factory II Introduction Portable DC Winding Resistance Tester is an innovative product with compact size hand held operation more portable and easy to carry The product is not only suita...

Page 5: ...tically selected which is convenient and fast Wide range and high accuracy 500u 50K Three phase automatic testing can be carried out and the three phase unbalance rate can be calculated automatically...

Page 6: ...and appearance Working power supply Built in lithium battery or external charger charger input 100 240VAC 50HZ 60HZ Charging voltage 16 8V Charging current 3A Charging time About 4 hr Use time More t...

Page 7: ...east 2 hours and continue to use the hand held tester when charging When the battery is fully charged the charger light changes from red to green VIII Tilt Hand held Tester In order to take the instru...

Page 8: ...5 Front view Side view Display panel Key Wristband...

Page 9: ...on the current operation Power switch The power switch of the whole machine is switched to the open position when it is switched on Switch to the close position when it is switched off Function module...

Page 10: ...icator of the printer are integrated When the printer is powered on the indicator light is normally on and it flashes when the paper is missing Press the button once and the printer passes the paper P...

Page 11: ...is low No Instruction 4 The main menu operation area of the instrument select corresponding functions through keys and press Enter to enter the corresponding menu DC resistance test function can measu...

Page 12: ...current temperature of the tested product and the temperature value is from 99 to 99 The device can monitor the oil temperature inside the test product in real time through an external wireless temper...

Page 13: ...t the appropriate function press the Enter button to execute the currently selected function and press the Cancel button to return to the previous screen No Instruction Retest according to the current...

Page 14: ...the power is not fully discharged It is not allowed to dismantle the test line and cut off the power switch during the test or discharge process In the process of testing transformers the unmeasured...