Withstanding Voltage lnstrument

Huazheng Electric Manufacturing (Baoding) Co., Ltd

Summary of Contents for CS2674AX

Page 1: ...CS2674AX Withstanding Voltage lnstrument Huazheng Electric Manufacturing Baoding Co Ltd ...

Page 2: ...ual and operate and maintain the instrument according to the manual to prolong its service life Just a light press the test will be completed automatically is the operating characteristics of this instrument If you are satisfied with this instrument please tell your colleagues if you are not satisfied with this instrument please call 0312 6775656 to tell you to serve you at all times Baoding Huazh...

Page 3: ...allation tester 2 II Notice Before Use 5 III Front Panel 8 IV Function outline 10 V Specifications 12 VI PLC interface 13 VII Operation Instructions and Procedures 14 VIII Applying Illustration 15 IX Precautions 16 X Packing list 17 ...

Page 4: ...er cord CS26016 6 high voltage test clip red CS26009 3 test clip black 3 Carton and packing materials Please save the original packing materials for use in future transportation 4 safety rules for using the instruments In the use of equipment be sure to follow the following safety rules 1 Do not use the tester in flammable gases environment In order to prevent the occurrence of an explosion or bur...

Page 5: ...strument overheating and damage Please do not accumulate other items in air inlet and outlet place to avoid blocking duct 6 Do not use the tester on the sloping surface or shaking of the places Instrument on a sloping surface or shaking of the place is easy to drop and break down 7 Do not use the tester beside the sensitive receiving equipment or testing equipment If the tester is near these equip...

Page 6: ...lothes decorated with metal or wear metal accessories such as watches metal bracelets Non cardiac patients or those who wear a pacemaker can not operate the tester 6 Check power supply voltage 1 Input supply voltage apparatus This instrument uses 120V AC 10 47 63Hz single phase power WARNING To prevent malfunction or damage to tester please use the tester in the specified voltage range 7 Check and...

Page 7: ...t to ground reliably 1 Connect the power cord into a grounded three phase power outlet If the three phase power supply plug is not grounded the instrument s rear panel has a protective earth terminal connect the protective earth ground terminal to safety ground Ground terminal on the rear panel as shown below II Notice Before Use Before operation please read the notice carefully Warning The maximu...

Page 8: ...e voltmeter used External voltage may damage the voltmeter 2 Emergency treatment In case of emergency electric shock or DUT burning must take the following two actions and the two actions must be all completed 1 Turn off the power switch 2 Unplug the power cord of the instrument 3 Precautions during testing 3 1 Wear insulated gloves against electric shock To prevent electric shock accidents in the...

Page 9: ...test probes and high voltage output are all with high voltage do not touch any part of the test loop Note When the host is high output state do not touch any part of the test circuit to avoid electric shock 3 6 Test Termination When the test has come to an end or without the use of tester or the tester is no longer used or user is left be sure set power switch as OFF That is to say turn off the po...

Page 10: ...and then start the tester rotate of the voltage regulating knob slowly to the required voltage value III Front Panel 一 Front panel illustration 二 Rear panel illustration 一 Front panel 1 Power switch It s used to control whether turn on off the power 2 Start Pressing it the test light illuminates and the instrument begins to work 3 Reset Pressing it the test light goes out No high voltage is output...

Page 11: ...es it shows current has output contrariwise the current has been cut off 13 Voltage displays AC DC 0 03 19 99kV 14 Leakage current alarm value display 0 3mA 10mA 15 Time displays 1 999s count down 16 UP key 17 DOWN key 18 Test pre set key 19 2mA 20mA switch gear 20 Voltage switching converter normal for AC pressing it for DC 21 Alarm value of leakage current adjustment button Press preset switch w...

Page 12: ...tage and current can be displayed in real time Pre set leak current can be set arbitrarily The time resolution is 0 1s time set range from 0 0s to 999s In addition it is equipped with PLC interface Standards IEC60065 Safety for audio video and similar electronic apparatus IEC60204 1 Safety of machinery IEC60335 1 Safety of household and similar electrical appliances IEC60598 1 Safety of lighting e...

Page 13: ... whether it is time testing or continuous testing users can stop the tester by pressing the STOP button or enable the reset signal 3 Remote interface PLC interface This hipot tester can be equipped with remote control gun or the PLC interface to start or reset 4 Danger indicator To ensure operator safety the danger indicator will be lit if there is high voltage After the hipot tester starts to tes...

Page 14: ...999s 0 0 continuous test D C W Output voltage range 0 00 20 00 kV Maximum output power 200VA Maximum rated current 10mA Current gear 2mA 10mA Ripple factor 5 Testing time 0 0s 999s 0 0 continuous test Voltmeter Range 0 00 20 00 kV Accuracy 5 3dgt Resolution 10V Display value Root mean square Ammeter Measuring range AC 0 100mA 10 00mA DC 0 100mA 10 00mA Resolution 2mA Gear 1uA 10mA Gear 10uA Measur...

Page 15: ...IN 4 Signal output under testing Between PIN 2 and PIN 5 Test fail signal Between PIN 6 and PIN 7 Test pass signal Between PIN 8 and PIN 9 3 Explain the Input Output Signal Connection This tester equips with remote control joints and the TEST and RESET function of this instrument can be manipulated by the outside remote control device These joints provide the power supply that holds control functi...

Page 16: ...ss the up button continuously and hold the button the time preset value will increase 1 continuously If a certain value is added to the time preset value will increase 10 until up to 999s Decrease Press the DOWN button on the front panel the time preset value will minus 1 Press the down button continuously and hold the button the time preset value will minus 1 continuously If a certain value is de...

Page 17: ...ion pad when operate the hipot tester Note the withstanding voltage of insulating gloves and insulated pad should be twice than the maximum output voltage of the tester 2 Set the alarm value of leak current and test time Please refer to the seventh chapter 3 Connect the DUT Device Under Test to the tester Please confirm the following before connection 1 The tester is shut down or on the reset stat...

Page 18: ...time is return to 0 and the tester is not alarm the test is passed and the DUT Device Under Test T is eligible Note During DCW test confirm the measurement circuit is without any electricity before replace another device IX Precautions 9 2 Maintenance 9 2 1Warranty 1 Maintenance period The instrument is warranted to be free from use for a period of 12 months from the date of shipment to the origin...

Page 19: ... case during the testing process the case will have a high voltage existance which is very dangerous So long as any contact of the case person may have resulted in the occurrence of electric shock that is why this ground terminal must be reliable connection with the earth 3 When the tester s power switch is turned on please do not touch the any items which is connected to the high voltage output p...