7
3.3 Connect the test leads to high voltage output
After connected the test line of current measurement, then connect high-voltage output
line obey the following procedure:
(1) Press [STOP] key.
(2) To confirm the test light does not light.
(3)
High-voltage output cable plugs into the high voltage output port.
3.4 Replace DUT
(4) When one test is completed, the replacement of another DUT, please confirm:
(5) (1) Tester is in the "reset" state.
(6) (2) Test light is not blinking, the discharge lamp is not lit.
(7) (3) LCD voltage display number is not beating.
Warning: Replace the DUT, do not touch high-voltage probe!
3.5 In testing status
When the tester is in testing status, test leads, DUT, test probes, and high-voltage output
are all with high voltage, do not touch any part of the test loop.
Note: When the host is high output state, do not touch any part of the test circuit to avoid
electric shock!
3.6 Test Termination
When the test has come to an end or without the use of tester or the tester is no longer
used, or user is left, be sure set power switch as OFF. That is to say, turn off the power
switch.
3.7 After the test, user must confirm the following:
At any time, before touch high-voltage leads, high voltage output or the DUT, make sure
that:
(1) Power switch is off, the displayer does not light.
(2) After IR test or DC test, the test body maybe still exist high-voltage after the power
switch is off, so it still needs some time to fully discharge. Even if test is finished, please
do not immediately touch any place where may cause electric shock.
3.8 Attention