I.    Safety rules and precautions 

Thank you for purchasing our company's Ground pile clamp earth resistance tester. In order to 
better use this product, please be sure to: 

——Read this user manual in detail. 
——Strictly follow the safety rules and precautions listed in this manual. 

 

²

 

This instrument according to IEC61010 safety specifications for design, production, 

inspection.   

²

 

In any case, use the meter should pay special attention to safety. 

²

 

When  measuring,  please  do  not  use  the  high  frequency  signal  generator  such  as  mobile     

phone near the instrument, so as not to cause error.   

²

 

Pay attention to the label and symbol of the instrument. 

²

 

Make sure the instrument and accessories are in good condition before use. 

²

 

Before starting up, press the trigger once or twice to make sure the jaws are well closed. 

²

 

Do not measure in flammable places, sparks may cause an explosion. 

²

 

When starting up, do not press the trigger or clamp any wires. 

²

 

Power  on,  the  “OL”  symbol  is  displayed  during  the  clamp  method  measurement,  then 

measured object can be clamped. 

²

 

Do not place or store the meter for a long time in a place with high temperature, humidity, 

condensation, or under direct sunlight. 

²

 

Must turned off the meter When replacing the battery. 

²

 

When the low battery low voltage symbol "

" display, the battery should be replaced in 

time, otherwise it will cause errors. 

²

 

The contact surface of the jaws must be kept clean and cannot be wiped with corrosive agents 

and rough objects. 

²

 

When opening the trigger, avoid impact on the clamp meter, especially the jaw joint surface. 

²

 

The clamp will make a slight noise when measuring resistance. This is normal.Note the "beep-

beep-beep-" sound to distinguish the alarm. 

²

 

Pay attention to the measurement range and use environment specified of this instrument. 

²

 

The measurement lead current should not exceed the upper limit of the clamp meter. 

²

 

Use,  dis  assembly,  calibration  and  maintenance  of  this  instrument  must  be  operated  by 

authorized personnel. 

²

 

Due  to  the  reasons  of  this  instrument,  if  it  is  dangerous  to  continue  using  it,  it  should  be 

stopped immediately, sealed up immediately, and handled by an authorized organization. 

²

 

The "          " safety warning signs in the instrument and manual must be strictly followed by 

the user Allow safe operation. 

 

II

  Introduction 

Ground pile clamp earth resistance tester also name Loop resistance tester,used for 

grounding resistance test. 

This instrument uses 2.4-inch color screen design, with clamp resistance, three or four wire 

resistance, AC voltage, current measurement functions, and also has data storage, data 
access, alarm, automatic shutdown and other functions. 

This meter is beautiful and high-grade, wide range, high resolution, convenient operation, 

easy to carry, accurate, reliable, stable performance, strong anti-interference ability.It also has a 

Summary of Contents for CE-8201

Page 1: ...MANUAL CE 8201...

Page 2: ...rement principle 8 8 4 Principle of Soil Resistivity Measurement 9 8 5 Working error principle 9 8 6 AC current measurement principle 10 8 7 Ground voltage measurement principle 10 IX Operation Method...

Page 3: ...ced in time otherwise it will cause errors The contact surface of the jaws must be kept clean and cannot be wiped with corrosive agents and rough objects When opening the trigger avoid impact on the c...

Page 4: ...rage and data upload functions III Range and accuracy Measurement mode Range Resolution Accuracy Clamp method 0 02 9 99 0 01 2 rdg 0 1 10 0 99 9 0 5 3 rdg 0 5 100 199 1 5 rdg 1 200 299 5 10 rdg 5 300...

Page 5: ...ance Range 0 00 1800 Leakage current range 0 000mA 50A Current resolution 1uA Three four wire method range 0 00 30k Soil resistivity range 0 00 9999k m Ground voltage range 600V Bluetooth Built in rec...

Page 6: ...e storage is full Overflow display OL symbol indication when over range overflow Alarm function When the measured value exceeds the alarm set value an alarm will be issued battery voltage Real time di...

Page 7: ...U key 4 5 Power button 6 TEST 7 8 Screen 9 E port 10 H port 11 USB interface 12 Charging indicator 13 ES port 14 S port VII Interface display 7 1 Interface 1 Main interface 2 Pincer 3 Three four line...

Page 8: ...time the accuracy of the measurement cannot be guaranteed and the battery should be replaced The OL symbol indicates that the measured value exceeds the measuring range of the clamp meter Alarm symbo...

Page 9: ...measures I and calculates the measured resistance R 8 2 Three four wire method measurement principle The three wire four wire method measures grounding resistance which is suitable for accurately meas...

Page 10: ...method The alternating current I flows between the E ground electrode and the H current electrode find the potential difference V between the S voltage electrode and the ES auxiliary ground electrode...

Page 11: ...EST key to enter 2 Check whether the meter displays OL then resistance measurement can be performed Otherwise you need to check whether the clamp head is tight 3 Open the clamp head and clamp into the...

Page 12: ...o enter 2 Open the clamp head and clamp the wire under test Wait for the value to stabilize and read the displayed current value 3 Short press the TEST key to keep the data and save it the screen disp...

Page 13: ...rt the test wire into the bottom port of the meter according to the color as shown in the figure below Wait for the value to stabilize and read the displayed resistance value 4 Press the TEST key to k...

Page 14: ...ort press the MENU key in the layer where the alarm type is selected to return to the menu interface 9 Long press the MENU button on any interface to enable or disable the alarm function 10 X Battery...

Page 15: ...technical sense but Since the grounding hemisphere of each tower is much smaller than the distance between the towers and after all the number of grounding points is large R0 is much smaller than R1 T...

Page 16: ...le point grounding However users can use a test wire and a ground electrode near the grounding system to artificially create a loop for testing The following introduces two methods for measuring singl...

Page 17: ...resistances In this way you can easily calculate each ground resistance value because of R1 RA RB R2 RB RC R3 RC RA So RA R1 R3 R2 2 This is the ground resistance value of the grounding body RA In or...

Page 18: ...17 XIII Accessories Meter 1 pcs Test line 4 pcs Simple test line 2 pcs USB communication line 1 pcs Monitoring software CD 1 pcs Instructions warranty 1 set Instrument box 1 pcs...

Page 19: ...sed as a reason for using the product for special purposes The company is not responsible for other losses caused by use The company reserves the right to modify the contents of the user manual If the...

Reviews: