HTC CE-8201 Manual Download Page 14

 

-  13  - 

9.9 Alarm Settings 

 

 

1)  In the menu selection, select the "alarm function" function, press the 

TEST

 key to enter 

2)  The alarm thresholds for four measurement functions can be set. 
3)  After entering the alarm function interface, you can switch the set alarm type by short 

pressing the ▲ key or the ▼ key. 

4)  After  confirming  the  alarm  type,  short  press  the  TEST  key  to  enter  the  alarm  state 

selection, and then short press the ▲ key or the ▼ key to switch between setting greater 
than alarm (>) or less than alarm (<). 

5)  After confirming the alarm status, you can short press the TEST key to enter the alarm 

threshold modification status, short press the ▲ key to decrease the step or short press 
the ▼ key to increase the step to modify the threshold. 

6)  Long press the ▲ key to move the threshold step cursor to the right, and long press the 

▼ key to move the threshold step cursor to the left. 

7)  Short press the TEST key again to save the current settings and exit. 
8)  Or short press the MENU key to return to the previous layer, short press the MENU key 

in the layer where the alarm type is selected to return to the menu interface. 

9)  Long press the MENU button on any interface to enable or disable the alarm function. 
10)  

X.    Battery Instructions 

1)  When  the  battery  voltage  is  too  low,  the  battery  symbol  "

"  is  displayed,  please 

charge it in time. 

2)  Low battery voltage affects measurement accuracy 
3)  When the boot screen flashes, the screen goes black. It may be that the battery power is 

not enough to boot. Please charge it fully before starting measurement. 

4)  The battery symbol "

" flashes to indicate that it is about to be shut down, and the 

test cannot be performed at this time. Please fully charge the battery and start the test 
again. 

 

XI.    Mobile APP Instructions 

1)  After installing the APP of "Ground pile clamp earth resistance tester. apk" in the CD, open 

the APP to search for a Bluetooth device named "CLAMP_METER" (This function can 
only  be  used  for  devices  with  Bluetooth),  click  it  and  wait  for  successful  connection 
(Bluetooth PIN code: 123456 is required for the first connection). 

2)  If the mobile phone is connected to the instrument through Bluetooth, the software on the 

Alarm type 

Alarm State 

Alarm threshold 

Threshold step cursor

 

Summary of Contents for CE-8201

Page 1: ...MANUAL CE 8201...

Page 2: ...rement principle 8 8 4 Principle of Soil Resistivity Measurement 9 8 5 Working error principle 9 8 6 AC current measurement principle 10 8 7 Ground voltage measurement principle 10 IX Operation Method...

Page 3: ...ced in time otherwise it will cause errors The contact surface of the jaws must be kept clean and cannot be wiped with corrosive agents and rough objects When opening the trigger avoid impact on the c...

Page 4: ...rage and data upload functions III Range and accuracy Measurement mode Range Resolution Accuracy Clamp method 0 02 9 99 0 01 2 rdg 0 1 10 0 99 9 0 5 3 rdg 0 5 100 199 1 5 rdg 1 200 299 5 10 rdg 5 300...

Page 5: ...ance Range 0 00 1800 Leakage current range 0 000mA 50A Current resolution 1uA Three four wire method range 0 00 30k Soil resistivity range 0 00 9999k m Ground voltage range 600V Bluetooth Built in rec...

Page 6: ...e storage is full Overflow display OL symbol indication when over range overflow Alarm function When the measured value exceeds the alarm set value an alarm will be issued battery voltage Real time di...

Page 7: ...U key 4 5 Power button 6 TEST 7 8 Screen 9 E port 10 H port 11 USB interface 12 Charging indicator 13 ES port 14 S port VII Interface display 7 1 Interface 1 Main interface 2 Pincer 3 Three four line...

Page 8: ...time the accuracy of the measurement cannot be guaranteed and the battery should be replaced The OL symbol indicates that the measured value exceeds the measuring range of the clamp meter Alarm symbo...

Page 9: ...measures I and calculates the measured resistance R 8 2 Three four wire method measurement principle The three wire four wire method measures grounding resistance which is suitable for accurately meas...

Page 10: ...method The alternating current I flows between the E ground electrode and the H current electrode find the potential difference V between the S voltage electrode and the ES auxiliary ground electrode...

Page 11: ...EST key to enter 2 Check whether the meter displays OL then resistance measurement can be performed Otherwise you need to check whether the clamp head is tight 3 Open the clamp head and clamp into the...

Page 12: ...o enter 2 Open the clamp head and clamp the wire under test Wait for the value to stabilize and read the displayed current value 3 Short press the TEST key to keep the data and save it the screen disp...

Page 13: ...rt the test wire into the bottom port of the meter according to the color as shown in the figure below Wait for the value to stabilize and read the displayed resistance value 4 Press the TEST key to k...

Page 14: ...ort press the MENU key in the layer where the alarm type is selected to return to the menu interface 9 Long press the MENU button on any interface to enable or disable the alarm function 10 X Battery...

Page 15: ...technical sense but Since the grounding hemisphere of each tower is much smaller than the distance between the towers and after all the number of grounding points is large R0 is much smaller than R1 T...

Page 16: ...le point grounding However users can use a test wire and a ground electrode near the grounding system to artificially create a loop for testing The following introduces two methods for measuring singl...

Page 17: ...resistances In this way you can easily calculate each ground resistance value because of R1 RA RB R2 RB RC R3 RC RA So RA R1 R3 R2 2 This is the ground resistance value of the grounding body RA In or...

Page 18: ...17 XIII Accessories Meter 1 pcs Test line 4 pcs Simple test line 2 pcs USB communication line 1 pcs Monitoring software CD 1 pcs Instructions warranty 1 set Instrument box 1 pcs...

Page 19: ...sed as a reason for using the product for special purposes The company is not responsible for other losses caused by use The company reserves the right to modify the contents of the user manual If the...