HT9020
EN - 42
7.1.1. Reference
guidelines
Safety: IEC/EN61010-1,
IEC/EN61010-2-032
EMC :
IEC/EN61326-1
Technical documentation: IEC/EN61187
Safety of measuring accessories: IEC/EN61010-31
Insulation: double
insulation
Pollution level:
2
Max height of use:
2000m
Measurement category:
CAT IV 600V / CAT III 1000V to gnd, max 1000V
between inputs
7.1.2. General
characteristics
Mechanical characteristics
Dimensions (L x W x H):
252 x 88 x 44mm (9 x 3 x 2 in)
Weight (batteries included):
approx 420g (15 ounces)
Jaw opening / Max cable size:
45mm (1.8 in)
Power supply
Battery type:
2 batteries x 1.5V LR 03 AAA
Battery life:
approx. 150 hours of use in “W ” position
Auto power OFF:
after 5 min of idleness (disabled)
Display
Characteristics: graphic
display 128x128 pixels
Sampling rate:
128 samples per period (base sampling)
Updating frequency:
1time/s
7.2. ENVIRONMENT
7.2.1. Environmental conditions for use
Reference calibration temperature: 23° ± 5 °C (73 ± 41°F)
Operating temperature:
0 ÷ 40 °C (32 ÷ 104°F)
Allowable relative humidity:
<80%HR
Storage temperature:
-10 ÷ 60°C (14 ÷ 140°F)
Storage humidity:
<70%HR
This instrument satisfies the requirements of Low Voltage Directive 2006/95/EC
(LVD) and of EMC Directive 2004/108/EC
This instrument satisfies the requirements of 2011/65/EU (RoHS) directive and
2012/19/EU (WEEE) directive
7.3. ACCESSORIES
PROVIDED
Pair of test leads
Pair of alligator clips
Carrying
bag
Batteries
ISO9000 calibration certificate
User
manual
Summary of Contents for HT9020