HT9020
EN - 3
1.2. DURING
USE
Please carefully read the following recommendations and instructions:
CAUTION
Failure to comply with the Caution notes and/or Instructions may damage
the instrument and/or its components or be a source of danger for the
operator.
Before activating the switch, remove the conductor from the clamp jaw or disconnect
the test leads from the circuit under test.
When the instrument is connected to the circuit, do not touch any unused terminal.
Keep your hands always under the hand protection. This protection is always located in
a suitable position to guarantee a correct safety distance from possible exposed or live
parts (see Fig. 3)
Avoid measuring resistance if external voltages are present. Even if the instrument is
protected, excessive voltage could cause a malfunction of the clamp.
During current measurement, any other current near the clamp may affect
measurement accuracy.
When measuring current, always put the conductor as close as possible to the middle
of the clamp jaw, to obtain the most accurate reading.
While measuring, if the value or the sign of the quantity under test remain unchanged,
check if the HOLD function is enabled.
1.3. AFTER
USE
When measurement is complete, switch
OFF
the instrument.
If the instrument is not to be used for a long time, remove the batteries
1.4.
DEFINITION OF MEASUREMENT (OVERVOLTAGE) CATEGORY
Standard IEC/EN61010-1: “Safety requirements for electrical equipment for measurement,
control and laboratory use, Part 1: General requirements” defines measurement category,
commonly called overvoltage category. In § 6.7.4: Measured circuits, circuits are divided
into the following measurement categories:
(OMISSIS)
Measurement category IV
is for measurements performed at the source of the low-
voltage installation.
Examples are electricity meters and measurements on primary overcurrent protection devices
and ripple control units.
Measurement category III
is for measurements performed on installations inside
buildings.
Examples are measurements on distribution boards, circuit breakers, wiring, including cables,
bus-bars, junction boxes, switches, socket-outlets in the fixed installation, and equipment for
industrial use and some other equipment, for example, stationary motors with permanent
connection to fixed installations.
Measurement category II
is for measurements performed on circuits directly
connected to the low-voltage installation.
Examples are measurements on household appliances, portable tools and similar equipment.
Measurement category I
is for measurements performed on circuits not directly
connected to MAINS.
Examples are measurements on circuits not derived from MAINS, and specially protected
(internal) MAINS-derived circuits. In the latter case, transient stresses are variable; for that
reason, the standard requires that the transient withstand capability of the equipment is made
known to the user.
Summary of Contents for HT9020
Page 1: ... Copyright HT ITALIA 2016 Release 1 02 01 03 2016 HT9020 User manual ...
Page 2: ......
Page 50: ......
Page 51: ......