HT9019
EN - 13
6. TECHNICAL SPECIFICATIONS
6.1. CHARACTERISTICS
Uncertainty is calculated as [% rdg + (number of dgt) x resolution]. It is referred to the
following reference conditions: 18°C
28°C (65°F
83°F) with RH < 75%
DC Voltage (Autorange)
Range Resolution Uncertainty
Input
impedance
Overload
protection
600.0mV 0.01mV
(1.0%rdg + 3dgt)
10M
1000VDC/ACrms
6.000V 0.001V
60.00V 0.01V
600.0V 0.1V
1000V 1V
The meter emits a continuous sound with VDC >1000V
AC Voltage TRMS (Autorange)
Range Resolution Uncertainty
Input
impedance
Bandwidh
Overload
protection
6.000V 0.001V
(1.0%rdg+4dgt)
(50
60Hz)
(3.5%rdg+5dgt)
(61
400Hz)
10M
50-400Hz 1000VDC/ACrms
60.00V 0.01V
600.0V 0.1V
1000V 1V
Integrated sensor for AC voltage detection: LED turn on for phase-earth voltage > 100V, 50/60Hz
The meter emits a continuous sound with VDC >750V
Resistance and Continuity test (Autorange)
Range Resolution Uncertainty
Buzzer
Overload
protection
600.0
0.1
(1.0%rdg+5dgt)
≤
60
600VDC/ACrms
6.000k
0.001k
60.00k
0.01k
600.0k
0.1k
6.000M
0.001M
60.00M
0.01M
(2.0%rdg.+10dgt)
Test current of continuity test: < 0.35mA
AC TRMS Current
Range Resolution Uncertainty
(*) Bandwidh
Overload
protection
60.00A 0.01A
(2.8%rdg+12dgt)
50
60Hz
1000AACrms
600.0A 0.1A
(2.8%rdg+8dgt)
1000A 1A
(3.0%rdg+8dgt)
60.00A 0.01A
(4.5% rdg+10dgt)
61
400Hz
600.0A 0.1A
(5.0%rdg+10dgt)
1000A 1A
PEAK function: response time <10ms
(*) Referred to cable inside to the center of clamp jaws. Position sensitivity:
2.0%rdg