![background image](http://html1.mh-extra.com/html/ht/400-series/400-series_user-manual_2178071068.webp)
400
Series
EN - 66
Environmental parameters (AUX)
Parameter Range Resolution
Transduced
signal
Accuracy
Temperature
-20.0
80.0°C
0.1°C
-20
+80mV
(2.0
rdg + 2dgt)
-4.0
176.0°F
0.1°F
-4
+176mV
Humidity
0.0
100.0% RH
0.1% RH
0
+100mV
DC voltage
(0.0
999.9mV)
0.1mV
(0.2
999.9mV)
Luminance
0.001
20.00Lux 0.001
0.02Lux
0
+100mV
0.1
2000Lux
0.1
2Lux
0
+100mV
1
20000Lux
0.1
2Lux
0
+100mV
Mains analisys (PWR)
Frequency measurement
Range [Hz]
Resolution [Hz]
Accuracy
47.0
63.0
0.1
(2.0
rdg + 2dgt)
Voltage range:
5.0
265.0
Current range:
0.005
1.2 x FS
Voltage measurement
Range [V]
Resolution [V]
Accuracy
5.0
265.0
0.1V
(0.5
rdg + 2dgt)
Max crest factor:
1,5
Frequency: 47.0
63.0 Hz
Voltage harmonics measurement
Range [V]
Resolution [V]
Order
Accuracy
0.0
265.0
0.1V
2
15
(2.0
rdg + 5dgt)
16
49
(5.0
rdg + 10dgt)
Frequency fundamental
47.0
63.0 Hz
Current measurement
Range [A]
Resolution [A]
Accuracy
0.005
1.2 x FS
See Table 2
(1.0
rdg + 2dgt)
Max crest factor:
3
Frequency: 47.0
63.0 Hz
Current harmonics measurement
Range [V]
Resolution [A]
Order
Accuracy
0.005
1.2 x FS
See Table 2
2
15
(2.0
rdg + 5dgt)
16
49
(5.0
rdg + 10dgt)
Frequency fundamental
47.0
63.0 Hz
Current fundamental
0.020 x FS
Full scale [A]
Resolution [A]
Full scale [A]
Resolution [A]
1 0.001
300
0.1
10 0.01
400
0.1
30 0.01
1000
1
100 0.1
2000
1
200 0.1
3000
1
Table 2: Clamp and instrument full scale and resolution values