![HT 400 Series User Manual Download Page 67](http://html1.mh-extra.com/html/ht/400-series/400-series_user-manual_2178071067.webp)
400
Series
EN - 65
Tripping current ( )
IdN [mA]
Type
Range IdN [mA]
Resolution [mA]
Accuracy
≤
10
AC
(0.5
1.1) IdN
0.1IdN -0%,
+10%rdg
A
(0.3
1.1) IdN
> 10
AC
(0.5
1.1) IdN
A
(0.3
1.1) IdN
RCD type:
AC (
), A (
), general and selective
Tripping time
resolution 1ms,
Accuracy
(2.0%rdg + 2dgt)
Global earth resistance without RCD’s tripping (Ra)
Range [
] Resolution
[
]
Accuracy
1
1999
1
(5.0
rdg + 3dgt)
RCD type:
AC (
), A (
), general and selective
Test current
< ½ IdN, accuracy: -10%, +0% IdN
Contact voltage Ut
range: 0
2Ut lim, resolution: 0.1V, accuracy: -0%, +(5%rdg + 3dgt)
Line/Loop fault impedance
Phase-neutral and phase-earth voltage range (110
240V)
10%
Phase-phase voltage range
(110
415V)
10%
Frequency 50Hz
0.5Hz, 60Hz
0.5Hz
TT and TN systems
Range [
] Resolution
[
]
Accuracy
0.01
19.99
0.01
(5.0
rdg + 3dgt)
20.0
199.9
0.1
200
1999 (phase-earth only)
1
Max peak test current at test voltage
3A @ 127V, 6A @ 230V, 10A @ 400V
IT systems (first fault current)
Range [mA]
Resolution [mA]
Accuracy
5
999
1
(5.0
rdg + 3dgt)
Contact voltage (ULIM)
25V, 50V
Global earth resistance (Ra)
Range [
] Resolution
[
]
Accuracy
0.01
19.99
0.01
(5.0
rdg + 1
)
20.0
199.9
0.1
200
1999 (phase-earth only)
1
Test current:
<15mA
Contact voltage (ULIM):
25V, 50V
Phase-neutral and phase-earth:
(110
240V)
10%
Frequency: 50Hz
0.5Hz, 60Hz
0.5Hz
Phas sequence (123)
Phase-neutral and phase-earth voltage range (110
240V)
10%
Frequency 50Hz
0.5Hz, 60Hz
0.5Hz
Leakage current (LEAK)
Range [mV]
Resolution [mV]
Accuracy
1
1200.0
0.1
(1.0
rdg + 2dgt)
Max crest factor:
3
Response time:
10ms
Frequency: 50Hz
0.5Hz, 60Hz
0.5Hz