252
HP E1415 Command Reference
[SENSe]
channels to return DC voltage.
Parameters
Comments
•
The <range> parameter: The HP E1415 has five ranges: .0625VDC, .25VDC,
1VDC, 4VDC, and 16VDC. To select a range, simply specify the range value
(for example, 4 selects the 4VDC range). If you specify a value larger than
one of the first four ranges, the HP E1415 selects the next higher range (for
example, 4.1 selects the 16VDC range). Specifying a value larger than 16
causes an error. Specifying 0 selects the lowest range (.0625VDC).
Specifying AUTO selects auto range. The default range (no range parameter
specified) is auto range.
•
If you are using amplifier SCPs, you should set them first and keep their
settings in mind when specifying a range setting. For instance, if your expected
signal voltage is to be approximately .1VDC and the amplifier SCP for that
channel has a gain of 8, you must set <range> no lower than 1VDC or an input
out-of-range condition will exist.
•
The *CAL? command calibrates channels based on Sense Amplifier SCP setup
at the time of execution. If SCP settings are changed, those channels are no
longer calibrated. *CAL? must be executed again.
•
See “Linking Input Channels to EU Conversion” on page 64.
•
When Accepted: Not while INITiated
•
Related Commands: *CAL?, INPUT:GAIN
…
•
*RST Condition: SENSE:FUNC:VOLT AUTO,(@100:163)
Usage
FUNC:VOLT (@140:163)
Channels 40 - 63 measure voltage in
auto-range (defaulted)
[SENSe:]REFerence
[SENSe:]REFerence <
type
>,<
sub_type
>,[<
range
>,](@<
ch_list
>)
links channel
in <ch_list> to the reference junction temperature EU conversion based on type and
sub_type. When scanned, the resultant value is stored in the Reference Temperature
Register, and by default the FIFO and CVT. This is a resistance temperature
measurement and uses the on-board 122
µ
A current source.
Note
The reference junction temperature value generated by scanning the reference
channel is stored in the Reference Temperature Register. This reference temperature
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
range
numeric (float32)
see comments
VDC
ch_list
channel list (string)
100 - 163
none
Summary of Contents for VXI 75000 C Series
Page 2: ......
Page 16: ...16 ...
Page 18: ......
Page 30: ...30 Getting Started Chapter 1 Notes ...
Page 32: ...32 Field Wiring Chapter 2 Figure 2 1 Channel Numbers at SCP Positions ...
Page 44: ...44 Field Wiring Chapter 2 Figure 2 11 HP E1415 Terminal Module ...
Page 54: ...54 Field Wiring Chapter 2 Notes ...
Page 61: ...Programming the HP E1415 for PID Control 61 Chapter 3 Programming Overview Diagram ...
Page 136: ...136 Creating and Running Custom Algorithms Chapter 4 Notes ...
Page 152: ...152 Algorithm Language Reference Chapter 5 Notes ...
Page 304: ...304 HP E1415 Command Reference Chapter 6 Command Quick Reference Notes ...
Page 308: ...308 Specifications Appendix A Thermocouple Type E 200 800C SCPs HP E1501 02 03 ...
Page 309: ...Specifications 309 Appendix A Thermocouple Type E 200 800C SCPs HP E1508 09 ...
Page 310: ...310 Specifications Appendix A Thermocouple Type E 0 800C SCPs HP E1501 02 03 ...
Page 311: ...Specifications 311 Appendix A Thermocouple Type E 0 800C SCPs HP E1509 09 ...
Page 312: ...312 Specifications Appendix A Thermocouple Type E Extended SCPs HP E1501 02 03 ...
Page 313: ...Specifications 313 Appendix A Thermocouple Type E Extended SCPs HP E1508 09 ...
Page 314: ...314 Specifications Appendix A Thermocouple Type J SCPs HP E1501 02 03 ...
Page 315: ...Specifications 315 Appendix A Thermocouple Type J SCPs HP E1508 09 ...
Page 316: ...316 Specifications Appendix A Thermocouple Type K SCPs HP E1501 02 03 ...
Page 317: ...Specifications 317 Appendix A Thermocouple Type R SCPs HP E1501 02 03 ...
Page 318: ...318 Specifications Appendix A Thermocouple Type R SCPs HP E1508 09 ...
Page 319: ...Specifications 319 Appendix A Thermocouple Type S SCPs HP E1501 02 03 ...
Page 320: ...320 Specifications Appendix A Thermocouple Type S SCPs HP E1508 09 ...
Page 321: ...Specifications 321 Appendix A Thermocouple Type T SCPs HP E1501 02 03 ...
Page 322: ...322 Specifications Appendix A Thermocouple Type T SCPs HP E1508 09 ...
Page 323: ...Specifications 323 Appendix A 5K Thermistor Reference SCPs HP E1501 02 03 ...
Page 324: ...324 Specifications Appendix A 5K Thermistor Reference SCPs HP E1508 09 ...
Page 325: ...Specifications 325 Appendix A RTD Reference SCPs HP E1501 02 03 ...
Page 326: ...326 Specifications Appendix A RTD SCPs HP E1501 02 03 ...
Page 327: ...Specifications 327 Appendix A RTD SCPs HP E1508 09 ...
Page 328: ...328 Specifications Appendix A 2250 Thermistor SCPs HP E1501 02 03 ...
Page 329: ...Specifications 329 Appendix A 2250 Thermistor SCPs HP E1508 09 ...
Page 330: ...330 Specifications Appendix A 5K Thermistor SCPs HP E1501 02 03 ...
Page 331: ...Specifications 331 Appendix A 5K Thermistor SCPs HP E1508 09 ...
Page 332: ...332 Specifications Appendix A 10K Thermistor SCPs HP E1501 02 03 ...
Page 333: ...Specifications 333 Appendix A 10K Thermistor SCPs HP E1508 09 ...
Page 334: ...334 Specifications Appendix A Notes ...
Page 346: ...346 Glossary Appendix C Notes ...
Page 388: ...388 Generating User Defined Functions Appendix F Notes ...