HP E1415 Command Reference
245
[SENSe]
[SENSe:]FUNCtion:CUSTom:TCouple
[SENSe:]FUNCtion:CUSTom:TCouple <
type
>,[<
range
>,](@<
ch_list
>)
links
channels with the custom Engineering Unit Conversion table loaded with the
DIAG:CUST:PIECE command. The table is assumed to be for a thermocouple and
the <type> parameter will specify the built-in compensation voltage table to be used
for reference junction temperature compensation. SENS:FUNC:CUST:TC allows
you to use an EU table that is custom matched to thermocouple wire you have
characterized. Contact your Hewlett-Packard System Engineer for more information
on Custom Engineering Unit Conversion for your application.
Parameters
Comments
•
See “Linking Input Channels to EU Conversion” on page 64..
•
The <range> parameter: The HP E1415 has five ranges: .0625VDC, .25VDC,
1VDC, 4VDC, and 16VDC. To select a range, simply specify the range value
(for example, 4 selects the 4VDC range). If you specify a value larger than
one of the first four ranges, the HP E1415 selects the next higher range (for
example, 4.1 selects the 16VDC range). Specifying a value larger than 16
generates an error. Specifying 0 selects the lowest range (.0625VDC).
Specifying AUTO selects auto range. The default range (no range parameter
specified) is auto range.
•
If you are using amplifier SCPs, you should set them first and keep their
settings in mind when specifying a range setting. For instance, if your expected
signal voltage is to be approximately .1VDC and the amplifier SCP for that
channel has a gain of 8, you must set <range> no lower than 1VDC or an input
out-of-range condition will exist.
•
The sub_type EEXTended applies to E type thermocouples at 800
°
C and
above.
•
The *CAL? command calibrates temperature channels based on Sense
Amplifier SCP setup at the time of execution. If SCP settings are changed,
those channels are no longer calibrated. *CAL? must be executed again.
•
Related Commands: DIAG:CUST:PIEC, *CAL?,SENS:REF, and
SENS:REF:TEMP
•
*RST Condition: all custom EU tables erased
Usage
program must put table constants into array table_block
DIAG:CUST:PIEC 1,table_block,(@100:107)
send characterized thermocouple table
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
type
discrete (string)
E | EEXT | J | K | N | R | S | T
none
range
numeric (float32)
see comments
VDC
ch_list
channel list (string)
100 - 163
none
Summary of Contents for VXI 75000 C Series
Page 2: ......
Page 16: ...16 ...
Page 18: ......
Page 30: ...30 Getting Started Chapter 1 Notes ...
Page 32: ...32 Field Wiring Chapter 2 Figure 2 1 Channel Numbers at SCP Positions ...
Page 44: ...44 Field Wiring Chapter 2 Figure 2 11 HP E1415 Terminal Module ...
Page 54: ...54 Field Wiring Chapter 2 Notes ...
Page 61: ...Programming the HP E1415 for PID Control 61 Chapter 3 Programming Overview Diagram ...
Page 136: ...136 Creating and Running Custom Algorithms Chapter 4 Notes ...
Page 152: ...152 Algorithm Language Reference Chapter 5 Notes ...
Page 304: ...304 HP E1415 Command Reference Chapter 6 Command Quick Reference Notes ...
Page 308: ...308 Specifications Appendix A Thermocouple Type E 200 800C SCPs HP E1501 02 03 ...
Page 309: ...Specifications 309 Appendix A Thermocouple Type E 200 800C SCPs HP E1508 09 ...
Page 310: ...310 Specifications Appendix A Thermocouple Type E 0 800C SCPs HP E1501 02 03 ...
Page 311: ...Specifications 311 Appendix A Thermocouple Type E 0 800C SCPs HP E1509 09 ...
Page 312: ...312 Specifications Appendix A Thermocouple Type E Extended SCPs HP E1501 02 03 ...
Page 313: ...Specifications 313 Appendix A Thermocouple Type E Extended SCPs HP E1508 09 ...
Page 314: ...314 Specifications Appendix A Thermocouple Type J SCPs HP E1501 02 03 ...
Page 315: ...Specifications 315 Appendix A Thermocouple Type J SCPs HP E1508 09 ...
Page 316: ...316 Specifications Appendix A Thermocouple Type K SCPs HP E1501 02 03 ...
Page 317: ...Specifications 317 Appendix A Thermocouple Type R SCPs HP E1501 02 03 ...
Page 318: ...318 Specifications Appendix A Thermocouple Type R SCPs HP E1508 09 ...
Page 319: ...Specifications 319 Appendix A Thermocouple Type S SCPs HP E1501 02 03 ...
Page 320: ...320 Specifications Appendix A Thermocouple Type S SCPs HP E1508 09 ...
Page 321: ...Specifications 321 Appendix A Thermocouple Type T SCPs HP E1501 02 03 ...
Page 322: ...322 Specifications Appendix A Thermocouple Type T SCPs HP E1508 09 ...
Page 323: ...Specifications 323 Appendix A 5K Thermistor Reference SCPs HP E1501 02 03 ...
Page 324: ...324 Specifications Appendix A 5K Thermistor Reference SCPs HP E1508 09 ...
Page 325: ...Specifications 325 Appendix A RTD Reference SCPs HP E1501 02 03 ...
Page 326: ...326 Specifications Appendix A RTD SCPs HP E1501 02 03 ...
Page 327: ...Specifications 327 Appendix A RTD SCPs HP E1508 09 ...
Page 328: ...328 Specifications Appendix A 2250 Thermistor SCPs HP E1501 02 03 ...
Page 329: ...Specifications 329 Appendix A 2250 Thermistor SCPs HP E1508 09 ...
Page 330: ...330 Specifications Appendix A 5K Thermistor SCPs HP E1501 02 03 ...
Page 331: ...Specifications 331 Appendix A 5K Thermistor SCPs HP E1508 09 ...
Page 332: ...332 Specifications Appendix A 10K Thermistor SCPs HP E1501 02 03 ...
Page 333: ...Specifications 333 Appendix A 10K Thermistor SCPs HP E1508 09 ...
Page 334: ...334 Specifications Appendix A Notes ...
Page 346: ...346 Glossary Appendix C Notes ...
Page 388: ...388 Generating User Defined Functions Appendix F Notes ...