HP E1415 Command Reference
189
CALibration
CALibration:STORe
CALibration:STORe <
type
>
stores the most recently measured calibration
constants into Flash Memory (Electrically Erasable Programmable Read Only
Memory). When type=ADC, the module stores its A/D calibration constants as well
as constants generated from *CAL?/CAL:SETup into Flash Memory. When
type=TARE, the module stores the most recently measured CAL:TARE channel
offsets into Flash Memory.
Note
The HP E1415’s Flash Memory has a finite lifetime of approximately ten
thousand write cycles (unlimited read cycles). While executing CAL:STOR
once every day would not exceed the lifetime of the Flash Memory for
approximately 27 years, an application that stored constants many times
each day would unnecessarily shorten the Flash Memory’s lifetime. See
Comments below.
Parameters
Comments
•
The Flash Memory Protect jumper (JM2201) must be set to the enable position
before executing this command (See “Disabling Flash Memory Access
(optional)” on page 25.).
•
Channel offsets are compensated by the CAL:TARE command even when not
stored in the Flash Memory. There is no need to use the CAL:STORE TARE
command for channels which are re-calibrated frequently.
•
When Accepted: Not while INITiated
•
Related Commands: CAL:VAL:RES, CAL:VAL:VOLT
•
*RST Condition: Stored calibration constants are unchanged
Usage
CAL:STORE ADC
Store cal constants in non-volatile
memory after A/D calibration
CAL:STORE TARE
Store channel offsets in non-volatile
memory after channel tare
Command
Sequence
Storing A/D cal constants
perform complete A/D calibration, then...
CAL:STORE ADC
Storing channel tare (offset) values
CAL:TARE <
ch_list
>
to correct channel offsets
CAL:STORE TARE
Optional depending on necessity of long
term storage
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
type
discrete (string)
ADC | TARE
none
Summary of Contents for VXI 75000 C Series
Page 2: ......
Page 16: ...16 ...
Page 18: ......
Page 30: ...30 Getting Started Chapter 1 Notes ...
Page 32: ...32 Field Wiring Chapter 2 Figure 2 1 Channel Numbers at SCP Positions ...
Page 44: ...44 Field Wiring Chapter 2 Figure 2 11 HP E1415 Terminal Module ...
Page 54: ...54 Field Wiring Chapter 2 Notes ...
Page 61: ...Programming the HP E1415 for PID Control 61 Chapter 3 Programming Overview Diagram ...
Page 136: ...136 Creating and Running Custom Algorithms Chapter 4 Notes ...
Page 152: ...152 Algorithm Language Reference Chapter 5 Notes ...
Page 304: ...304 HP E1415 Command Reference Chapter 6 Command Quick Reference Notes ...
Page 308: ...308 Specifications Appendix A Thermocouple Type E 200 800C SCPs HP E1501 02 03 ...
Page 309: ...Specifications 309 Appendix A Thermocouple Type E 200 800C SCPs HP E1508 09 ...
Page 310: ...310 Specifications Appendix A Thermocouple Type E 0 800C SCPs HP E1501 02 03 ...
Page 311: ...Specifications 311 Appendix A Thermocouple Type E 0 800C SCPs HP E1509 09 ...
Page 312: ...312 Specifications Appendix A Thermocouple Type E Extended SCPs HP E1501 02 03 ...
Page 313: ...Specifications 313 Appendix A Thermocouple Type E Extended SCPs HP E1508 09 ...
Page 314: ...314 Specifications Appendix A Thermocouple Type J SCPs HP E1501 02 03 ...
Page 315: ...Specifications 315 Appendix A Thermocouple Type J SCPs HP E1508 09 ...
Page 316: ...316 Specifications Appendix A Thermocouple Type K SCPs HP E1501 02 03 ...
Page 317: ...Specifications 317 Appendix A Thermocouple Type R SCPs HP E1501 02 03 ...
Page 318: ...318 Specifications Appendix A Thermocouple Type R SCPs HP E1508 09 ...
Page 319: ...Specifications 319 Appendix A Thermocouple Type S SCPs HP E1501 02 03 ...
Page 320: ...320 Specifications Appendix A Thermocouple Type S SCPs HP E1508 09 ...
Page 321: ...Specifications 321 Appendix A Thermocouple Type T SCPs HP E1501 02 03 ...
Page 322: ...322 Specifications Appendix A Thermocouple Type T SCPs HP E1508 09 ...
Page 323: ...Specifications 323 Appendix A 5K Thermistor Reference SCPs HP E1501 02 03 ...
Page 324: ...324 Specifications Appendix A 5K Thermistor Reference SCPs HP E1508 09 ...
Page 325: ...Specifications 325 Appendix A RTD Reference SCPs HP E1501 02 03 ...
Page 326: ...326 Specifications Appendix A RTD SCPs HP E1501 02 03 ...
Page 327: ...Specifications 327 Appendix A RTD SCPs HP E1508 09 ...
Page 328: ...328 Specifications Appendix A 2250 Thermistor SCPs HP E1501 02 03 ...
Page 329: ...Specifications 329 Appendix A 2250 Thermistor SCPs HP E1508 09 ...
Page 330: ...330 Specifications Appendix A 5K Thermistor SCPs HP E1501 02 03 ...
Page 331: ...Specifications 331 Appendix A 5K Thermistor SCPs HP E1508 09 ...
Page 332: ...332 Specifications Appendix A 10K Thermistor SCPs HP E1501 02 03 ...
Page 333: ...Specifications 333 Appendix A 10K Thermistor SCPs HP E1508 09 ...
Page 334: ...334 Specifications Appendix A Notes ...
Page 346: ...346 Glossary Appendix C Notes ...
Page 388: ...388 Generating User Defined Functions Appendix F Notes ...