Installation Manual
CA294-03070 rev 00
page 15 of 153
Heading
Step
Reference
Procedure
Notes and
check
h. Click
Next
and
.
The color adjustment is
performed.
Make sure that the wizard
completes all of its steps, and that
all colors pass the color adjust
procedure.
i. Click
Finish
.
j. Load the desired substrate on the
press.
k. Print the CATP job.
l. In the 9th and 10th copy of the
CATP printout, measure the solid
patches (J area), and calculate
the average values.
Note the values.
m. Change the press software to
technician mode.
n. From the
Options
menu, click
Systab Editor
.
The Systab editor window opens.
o. Expand the
mmiCATopic
tree.
p. In the right window, double-click
SpecOdLowerDeltaOsm
.
The
mmiCATopic
SpecOdLowerDeltaOsm window
opens.
q. In the
Saved
field, type
-0.7
.
r. Click
OK
. The window closes.
s. Double-click
SpecOdLowerDelta5
.
The
mmiCATopic
SpecOdLowerDelta5 window
opens.
t. In
the
Saved
field, type
-0.7
.
u. Click
OK
. The window closes.
v. Close the Systab Editor.
w. From the
Adjustments
menu,
click
Color Adjustment &
Calibration
.
The Color Adjustment Wizard
Manager window opens.
x. In the
Automatic Color
Adjustment
area, click
Custom
Color Adjustment
.
The Custom Color Adjustment
wizard opens.
y. In the
Electrode Voltage
Calibration
field, make sure that
Include Velectrode Calibration is
not selected.
z. Click
Next
. The OD step opens.