2-8
Chapter 2
Specifications and Characteristics
Connector Pin Depth Information
Connector Pin Depth Information
1
1. Hewlett-Packard does not supply pin depth gauges for type-F or 7-16 connectors. The type-F mod-
ule adapter pin depth is set at –25.4 to –50.8 micrometers. The 7-16 module adapter pin depth is
set at 0 to –38.1 micrometers recessed from the nominal 1.77 mm offset between inner conductor
and outer conductor mating planes.
Table 2-3
3.5 mm Connector Pin Depth Limits
Device
Typical Pin Depth in
micrometers (10
–4
inches)
Measurement Uncertainty
1
in micrometers (10
–4
inches)
Observed Pin Depth Limits
in micrometers (10
–4
inches)
3.5 mm ECal Module
–25.4 to –50.8 (–10.0 to –20.0)
+1.3 to –1.3 (+0.5 to –0.5)
–24.1 to –52.1 (–9.5 to –20.5)
3.5 mm Adapter
–2.5 to –13.0 (–1.0 to –5.0)
+1.3 to –1.3 (+0.5 to –0.5)
–1.2 to –14.3 (–0.5 to –5.5)
1. Approxi2 sigma to –2 sigma of gauge uncertainty based on studies performed at the factory using analog gauges
according to recommended procedures.
Table 2-4
7 mm Connector Pin Depth Limits
Device
Typical Pin Depth in
micrometers (10
–4
inches)
Measurement Uncertainty
1
in micrometers (10
–4
inches)
Observed Pin Depth Limits
in micrometers (10
–4
inches)
7 mm ECal Module
–25.4 to –50.8 (–10.0 to –20.0)
+1.3 to –1.3 (+0.5 to –0.5)
–24.1 to –52.1 (–9.5 to –20.5)
1. Approxi2 sigma to –2 sigma of gauge uncertainty based on studies performed at the factory using analog gauges
according to recommended procedures.
Table 2-5
Type-N 50
Ω
Connector Pin Depth Limits
Device
Typical Pin Depth
1
in
micrometers (10
–4
inches)
Measurement Uncertainty
2
in micrometers (10
–4
inches)
Observed Pin Depth Limits
in micrometers (10
–4
inches)
Type-N 50
Ω
ECal Module –25.4 to –50.8 (–10.0 to –20.0) +3.8 to –3.8 (+1.5 to –1.5)
–21.6 to –54.6 (–8.5 to –21.5)
Type-N 50
Ω
Adapter
0 to –12.7 (0 to –5.0)
+3.8 to –3.8 (+1.5 to –1.5)
+3.8 to –16.5 (+1.5 to –6.5)
1. Measured from a 5.258 mm (0.207 inches) nominal offset.
2. Approxi2 sigma to –2 sigma of gauge uncertainty based on studies performed at the factory using analog gauges