5
Adjust the adjustment screws until they are making contact with the
interposer board. Ensure that the probe adapter is firmly seated in
the 190-pin Mictor connector, and that the probe adapter is parallel
within
±
4
°
with the interposer board (see illustration below).
C A U T I O N
The adjustment screws must be properly adjusted so that they are touching
the interposer board, and the probe adapter is parallel within
±
4
°
with the
interposer board. If the probe adapter is not supported by the adjustment
screws, or if the probe adapter is not seated so that it is parallel within
±
4
°
to the interposer board, damage may occur to the Mictor connectors.
Required alignment for connecting probe adapter to Intel’s Instrumented Interposer board
6
Attach the fully-connected probe adapter/Instrumented Interposer to the
connectors on the mother board.
Connecting to the Instrumented Interposer target system
To connect to the target system
10