Logic Analyzer Messages
This section lists some of the messages that the analyzer displays
when it encounters a problem.
“. . . Inverse Assembler Not Found”
This error occurs if you rename or delete the Pentium II processor
transaction tracker/inverse assembler that is attached to the configuration
file. Ensure that the transaction tracker/inverse assembler file is not renamed
or deleted.
“Measurement Initialization Error”
This error occurs when you have installed the cables incorrectly for the
HP 16550B cards. Ensure that your cable connections match the silk
screening on the card. Then, repeat the measurement.
See Also
The
HP 16550B Logic Analyzer Service Guide
.
“No Configuration File Loaded”
This is usually caused by trying to load a configuration file for one type of
module/system into a different type of module/system.
Verify that the appropriate module has been selected from the Load
{module} from File {filename} in the HP 16500B/C disk operation menu.
Selecting Load {All} will cause incorrect operation when loading most
preprocessor interface configuration files.
See Also
Chapter 1 describes how to load configuration files.
A–12
Preprocessor Interface for the Pentium II Processor