HP D300 Digital Dispenser
35
Figure 5-4
The difference between Global View and the well plate image
5-4 Set concentrations or volumes in wells
1.
Click and drag in the well plate image to select the wells where a fluid will be dispensed.
2.
To add the same single concentration or volume of fluid to the selected wells, right-click
anywhere in the well plate image and then click
Set Concentration
or
Set Volume
. The Set
Concentration or Set Volume dialog box appears. See
. Whether Set Concentration
or Set Volume is displayed is determined by the setting chosen in the Edit: Units pane of the
Options dialog box as described in
○
Select a fluid from the Fluid drop-down list box and then type in a concentration or volume
that is within the valid range that is stated in the dialog box. See
, shown for
concentration mode.
Figure 5-5
Set Concentration dialog box
NOTE:
The lowest possible concentration or volume is determined by the single drop
volume, the fluid concentration, and the assay volume. The maximum concentration or
volume is determined by the DMSO limit and the fluid concentration.
Valid concentration range
(from one small drop to the DMSO limit)
Edit check box
Global View
Well plate image