![HP 85325A Operating And Service Manual Download Page 84](http://html.mh-extra.com/html/hp/85325a/85325a_operating-and-service-manual_161508084.webp)
6-14
HP 85325A Operating and Service Manual
Service
Replaceable Parts
Miscellaneous Parts
The following parts are supplied with each system.
Adjustments
Only the HP 85309A LO/IF unit requires periodic circuitry adjustments.
Refer to the HP 85310A or HP 85301B Operating and Service Manual for
HP 85309A adjustment procedures. If the LO and RF source's timebases are
not connected together, refer to either of the manuals (mentioned above) for
adjustment instructions.
The LO power to the test and reference modules must be adjusted during
installation. Refer to Installation, Chapter 1 for more instructions.
Table 6-6
Miscellaneous Parts
Description
HP Part Number
Screws, captive
1390-0765
Screws, 1/2 inch
3030-0209
Screws, 3/8 inch
3030-0221
Nut, hex 4-40
2260-0002
Nut, hex 4-40
2260-0009
1
RF cable, SMA (m) / (m), 1 m
5061-5458
Wrench, 1/4 inch
8710-0510
Wrench, 3/16 inch
8720-0013
Torx driver T10
8710-1623
Modified wrench, 5/16
2
08555-20097
Attenuator, fixed, 10 dB, 3.5 mm
HP 8493C, opt. 010
Operating and Service Manual
85325-90001
1. Only supplied with R band.
2. This wrench is specially designed. The “closed” end has been
machined to be smaller, and has been slotted so it will fit over a cable.
The modified end has another feature as well, it slips when
approximately 10 inch-pounds of force is exerted on the SMA
connection. 10 inch-pounds is the torque specification for all SMA
connections inside the mm-wave modules, or inside the HP 85309A.