HP 85325A Operating and Service Manual 4-5
System Performance Verification
Receiver Performance Verification
Receiver Performance Verification
The receiver performance verification checks the receiver section of the
antenna measurement system. The tests are divided into two sections.
First Section
The first section tests the various parameters of the receiver system. The
results of these tests are used to calculate the specified parameters in the
section below. It is possible to have one of these tests fail but still be within
the specified parameters of the millimeter subsystem. These tests consist of:
RF to IF Conversion
Measures the power loss of the mixers when the RF signal is converted to
the IF signal. This is not a specified system performance parameter.
IF Noise
Measures sensitivity (the RMS noise level of the trace data when the test or
reference channels are terminated with a 50 ohm load). This is not a
specified system performance parameter.
Crosstalk
Measures signal leakage between the antenna reference and the test
channels. This is a specified system performance parameter.
Compression
Measures input power at which the mixers exhibit 0.1 dB compression. This
is a specified system performance parameter.
Second Section
The second section takes the results of the above section and calculates the
specified parameters of the millimeter subsystem. These tests consist of:
Tracking Results
Calculates the conversion gain tracking between channels using RF to IF
conversion measurements from the first section.This is not a specified
system performance parameter.
Sensitivity Results
Calculates the actual sensitivity of the receiver in RMS power using IF
Noise and RF to IF Conversion measurements from the first section.This is a
specified system performance parameter.
Crosstalk Results
Calculates the actual crosstalk between all channels using Crosstalk and RF
to IF Conversion measurements from the first section. This is a specified
system performance parameter.