4 wires. Jump over the third phase measurement
(5) This feature file, no maximum and minimum measurement
Warning
To avoid damages to the meter or harms to you,do you measure
higher than AC voltage 600(r.m.s )and AC Current 1000A (r.m.s)
Frequency measurement is the voltage signal measured the frequency, not
the current signal of frequency
。
I Active Energy(kWh)+Active Power(kW)+Time(mm:s)
(1) Set the Rotary Switch to "
kWh
" position.
(2) Clamp the corresponding target test lead ( Clamp A test lead for
measure A phase.... )
(3) The connecting method of 3 phases 4 wires, 3 phases 3 wires or single
phase 2 wires, see figure 4, 5 and 6
(4)Press
SELECT
to choose one of
A
,
B
or
C
phase (as figure 30)
● The double display shows tested object’s active engergy kWh value and
the measuring time of the corresponding phase.
● The measuring reading gets increasing along with the time increases.
Press
HOLD
to read a particular time kWh value. Then the reading and time
are locked, but still continuous accumulate measuring time.
● After read the data, press
HOLD
again to continous measurement. kWh
value continous accumulate.
● When the measuring time is over 24 hours or the Meter is switched to
other measuring ranges, active energy measuring will stop.
● The maximum reading of active energy is 9999kWh.
9999
will be
displayed when the reading is over than that.
(5) This feature file, no maximum and minimum measurement
(6) Press CLEAR to reset the time
(7) No USB communication