25
The double display shows the
φA value of phase Angle φ,Voltage value and
also Current value.
Then press SELECT to choose Second phase φB
figure 25
The double
display shows the φB value of phase Angle φ,Voltage value and
also Current value.
Then press SELECT to choose Second phase φC
figure
26
Measuring 1 phase and 2 phase as same as 3 phase 4 wires.
Skip the third
phase measurement
(5) This feature file, no maximum and minimum measurement
Warning
To avoid damages to the meter or harms to you,do you measure
higher than AC voltage 600(r.m.s )and AC Current 1000A (r.m.s)
H. Frequency(Hz)+Voltage(V)+Current(A)
(1)
Set the Rotary Switch to "
Hz
" position.
(2) Press the lever to open the transformer jaw, and clamp them to the
corresponding phase of tested conductor. If user needs to measure any
phase of the 3 phase, then clamp them to that phase’s conductor.