HP 85027A/B/C Service
ALLOWABLE RANGE (ALL CONNECTORS):
+0.002 INCH TO +0.010 INCH
CENTER COLLET MUST PROTRUDE BEYOND PLANE OF OUTER CONDUCTOR 0.002 —0.010 INCH (0.05
—
0.25mm).
ALLOWABLE RANGE
(ALL CONNECTORS):
+0.002 INCH TO
+0.010 INCH
OUTER CONDUCTOR
PLANE OF
OUTER
CONDUCTOR
CENTER COLLET
(PROTRUSION: +0.002 TO +0.010
INCH)
CENTER CONDUCTOR
(RECESSION:
—0.002 TO —0.008
INCH)
Figure 8-10. Gaging A PC-7 Connectors
8-17
PLANE OF
OUTER CONDUCTOR
PROTRUSION OF
CENTER COLLET
(+0.002
—
0.010 INCH)
DO NOT USE
ALIGNING PIN.