HP 85027A/B/C Adjustments
AC Adjustment Procedure
1.
Set up the equipment as shown in Figure 5-1. Turn on the analyzer, source and
power meter and allow 30 minutes for warm-up. Refer to Figure 5-2 for
locations of the adjustments.
2.
Connect the power meter sensor to the calibrated 10 dB step attenuator.
NOTE
If your attenuator does not have calibration data, determine
exactly how much the attenuation changes between the 0dB
and 30dB settings. Determine this with the source set for
about +13 dBm at 50 MHz CW. Use this figure when the
procedure refers to calibrated 30 dB; for example, 29.9 dB or
30.06 dB.
3.
Set the attenuator for 0 dB.
4.
Preset the analyzer. Set the source to generate 50 MHz CW without modulation.
Adjust the output for a reading of +6.5 dBm on the power meter and then turn on
the modulation.
Figure 5-2. Locations of A djustment Potentiometers
5-2