3-7
Calibration Overview
User Calibrations—Optical and Electrical
User Calibrations—Optical and Electrical
The following calibrations can be performed by the user:
O/E User Wavelength Calibration
Plug-in Module Vertical Calibration
Offset Zero Calibration
Dark Calibration
Probe Calibration
Channel Skew
External Scale
Electrical channels have calibration procedures for:
• adjusting timebase skew, for matching propagation delay between channels,
probes, cables, and so forth
• using external probes
Optical channels have calibration procedures for:
• adjusting timebase skew
• monitoring and adjusting internal offsets
• performing a user-defined O/E responsivity adjustment
C A U T I O N
The input circuits can be damaged by electrostatic discharge (ESD). Avoid
applying static discharges to the front panel input connectors. Before
connecting a coaxial cable to the connectors, momentarily short the center and
outer connectors of the cable together. Avoid touching the front panel input
connectors without first touching the frame of the instrument. Be sure the
instrument is properly earth-grounded to prevent buildup of static charge. An
antistatic mat and wristband are strongly recommended, particularly when
working with TDR modules.
Summary of Contents for 83487A
Page 1: ...User s Guide HP 83487A Optical Electrical Plug In Module ...
Page 6: ...vi General Safety Considerations ...
Page 8: ...Contents 2 Contents ...
Page 23: ...2 Channel Setup Menu 2 2 Displaying the Channel Setup Menus 2 5 Channel Setup Menu ...
Page 25: ...2 3 Channel Setup Menu Channel Setup Menu Figure 2 1 Optical Channel Setup menu ...
Page 26: ...2 4 Channel Setup Menu Channel Setup Menu Figure 2 2 Electrical Channel Setup menu ...
Page 58: ...3 22 Calibration Overview Complete Calibration Summary ...
Page 68: ...4 10 Specifications and Regulatory Information Declaration of Conformity ...
Page 98: ...Index 4 Index ...