2-13
Channel Setup Menu
Calibrate
Cal status
The
Cal status
function displays a screen similar to
.
Key Path
Channel,
Calibrate,
Cal Status
Figure 2-3. A typical Cal Status display.
Current Date
This is the current date and time. You can compare this to the last plug-in
module calibration time to see how long it has been since calibration was per-
formed.
Current Frame
∆
Temp
This is the temperature change on the inside of the instrument since the last
mainframe calibration was performed. The number indicates how many
degrees warmer or cooler the mainframe is compared to the last calibration.
Channel 1
Calibration
Status
The instrument displays
Calibrated
or
Uncalibrated
, depending on
whether the last plug-in module calibration is still valid. A calibration can be
invalidated if:
• The mainframe has cycled power.
• The plug-in has been repaired, reprogrammed, or removed from the
mainframe.
• The instrument’s operating temperature has changed and remains more than
5
°
C from the temperature at which the Plug-in calibration was performed.
Uncalibrated
indicates the plug-in module vertical calibration is invalid.
Summary of Contents for 83487A
Page 1: ...User s Guide HP 83487A Optical Electrical Plug In Module ...
Page 6: ...vi General Safety Considerations ...
Page 8: ...Contents 2 Contents ...
Page 23: ...2 Channel Setup Menu 2 2 Displaying the Channel Setup Menus 2 5 Channel Setup Menu ...
Page 25: ...2 3 Channel Setup Menu Channel Setup Menu Figure 2 1 Optical Channel Setup menu ...
Page 26: ...2 4 Channel Setup Menu Channel Setup Menu Figure 2 2 Electrical Channel Setup menu ...
Page 58: ...3 22 Calibration Overview Complete Calibration Summary ...
Page 68: ...4 10 Specifications and Regulatory Information Declaration of Conformity ...
Page 98: ...Index 4 Index ...