3-5
Calibration Overview
Factory Calibrations
drawing that shows each switch’s function and protected position. Refer to the
optional
HP 83480A, 54750A Service Guide
for more details about the main-
frame calibration, and the position of the rear-panel memory protect switches.
C A U T I O N
To prevent access to the mainframe calibration switch, place a sticker over the
access hole to this switch.
C A U T I O N
Do not attempt a Mainframe calibration without consulting the
HP 83480A,
54750A Service Guide
.
A mainframe calibration should be performed on a periodic basis, annually, or
when the ambient operating temperature has changed by and remains 5
°
C dif-
ferent than the operating temperature at which the last mainframe calibration
was performed. To see how much the operating temperature has changed
since the last mainframe calibration and the date of the last mainframe calibra-
tion, check the Calibration status by pressing the following key sequence:
Utility
,
Calibrate
, and then
Cal status
on.
The temperature change is displayed at the top of the display as shown in the
following figure.
Figure 3-1. Current Frame
∆∆∆∆
Temp condition
Summary of Contents for 83487A
Page 1: ...User s Guide HP 83487A Optical Electrical Plug In Module ...
Page 6: ...vi General Safety Considerations ...
Page 8: ...Contents 2 Contents ...
Page 23: ...2 Channel Setup Menu 2 2 Displaying the Channel Setup Menus 2 5 Channel Setup Menu ...
Page 25: ...2 3 Channel Setup Menu Channel Setup Menu Figure 2 1 Optical Channel Setup menu ...
Page 26: ...2 4 Channel Setup Menu Channel Setup Menu Figure 2 2 Electrical Channel Setup menu ...
Page 58: ...3 22 Calibration Overview Complete Calibration Summary ...
Page 68: ...4 10 Specifications and Regulatory Information Declaration of Conformity ...
Page 98: ...Index 4 Index ...