2-8
Channel Setup Menu
Bandwidth/Wavelength....
Bandwidth/Wavelength....
You can use the
Bandwidth/Wavelength
....
softkey to change the bandwidth and
wavelength settings.
Bandwidth
This function is available on the electrical channel only.
You can use the
Bandwidth
function to select either the 12.4 GHz or 20 GHz
bandwidth.
Key Path
Channel,
Bandwidth/Wavelength
....
, Bandwidth
Filter
The
Filter
function turns the filter on and off.
Key Path
Channel
,
Bandwidth/Wavelength....
,
Filter On Off
Wavelength
This function is available on the optical channel only.
The
Wavelength
function selects the desired wavelength for calibrated mea-
surements. The factory calibrated wavelength is 850 nm. A user-calibrated
wavelength is also available and can be calibrated in the range from 750 nm to
860 nm. Refer to
Chapter 3, “Calibration Overview”
for additional information
on performing a calibration.
Key Path
Channel
,
Bandwidth/Wavelength
....
, Wavelength
Filter
This function selects the specific filter for the type of compliance testing to be
performed.
Key Path
Channel
,
Bandwidth/Wavelength
....,
Filter
,
1063 Mb/s
or
1250 Mb/s
Summary of Contents for 83487A
Page 1: ...User s Guide HP 83487A Optical Electrical Plug In Module ...
Page 6: ...vi General Safety Considerations ...
Page 8: ...Contents 2 Contents ...
Page 23: ...2 Channel Setup Menu 2 2 Displaying the Channel Setup Menus 2 5 Channel Setup Menu ...
Page 25: ...2 3 Channel Setup Menu Channel Setup Menu Figure 2 1 Optical Channel Setup menu ...
Page 26: ...2 4 Channel Setup Menu Channel Setup Menu Figure 2 2 Electrical Channel Setup menu ...
Page 58: ...3 22 Calibration Overview Complete Calibration Summary ...
Page 68: ...4 10 Specifications and Regulatory Information Declaration of Conformity ...
Page 98: ...Index 4 Index ...