Troubleshooting
1RRUORZ3)'7'VLJQDOEXWUHDJHQWLRQVDUHQRUPDO
83
No or low PFDTD signal,
but reagent ions are normal
Possible Cause
You are flowing any reagent gas but methane in PCI.
Action
Switch to methane.
Possible Cause
Wrong or corrupted tune file loaded.
Action
Check your tune file.
Possible Cause
No PFDTD in the calibrant vial.
Action
Inspect the calibration vial on the back of the flow controller. If necessary, fill the
vial with PFDTD. Do not fill the vial completely; keep the level at least 0.5 cm from
the top of the vial.
Possible Cause
The pressure of the methane entering the flow controller is too high.
Action
Make sure the regulator on the methane supply is set to 10 psig (70 kPa).
Possible Cause
The CI ion source is dirty.
Action
Clean the ion source. See the
Maintenance
chapter in this manual for more
information.
Possible Cause
The calibration valve was not purged after the vial was refilled.
Action
Purge the calibration valve as described in the
Maintenance
chapter. Then clean
the ion source.
Possible Cause
The calibrant vial was overfilled. Excess PFDTD can quench the chemical ionization
reactions.
Action
Check the level of the PFDTD in the calibration vial as described in
Maintenance
chapter.
Summary of Contents for 5973 MSD
Page 1: ...H Hardware Manual HP 5973 MSD PCI NCI ...
Page 6: ...6 ...
Page 7: ...1 General Information Specifications 9 Reagent gases and regulators 10 ...
Page 13: ... Hardware Reference 13 ...
Page 25: ... Chemical Ionization Theory 3RVLWLYH WKHRU 25 ...
Page 31: ... Chemical Ionization Theory 1HJDWLYH WKHRU 31 ...
Page 34: ...34 ...
Page 92: ...92 ...
Page 114: ...114 ...
Page 120: ...120 ...
Page 122: ...H Manual Part Number G1999 90001 Copyright 1998 Hewlett Packard Printed in USA 2 98 ...