Troubleshooting
3UHVVXUHGRHVQRWFKDQJHZKHQUHDJHQWIORZLVFKDQJHG
77
Pressure does not change when reagent flow is changed
Possible Cause
The reagent gas regulator is closed.
Action
Check and, if necessary, open the reagent gas regulator.
Possible Cause
The reagent gas regulator is set to the wrong pressure.
Action
Set the reagent gas regulator to 10 psi (70 kPa) for methane or to 3 – 10 psi
(20 – 70 kPa) for isobutane or ammonia.
Possible Cause
The valve on the reagent gas bottle is closed.
Action
Check and, if necessary, open the valve on the reagent gas bottle.
Possible Cause
The reagent gas supply is empty.
Action
Check, and if necessary, replace the reagent gas supply.
Possible Cause
Reagent lines kinked, bent, pinched, or disconnected.
Action
Inspect the reagent lines and repair any defects. Check especially to make sure the
reagent line is connected to the tee connector on the end of the GC interface.
Possible Cause
GC/MSD interface clogged or damaged.
Action
Check for flow and repair or replace components as indicated.
Summary of Contents for 5973 MSD
Page 1: ...H Hardware Manual HP 5973 MSD PCI NCI ...
Page 6: ...6 ...
Page 7: ...1 General Information Specifications 9 Reagent gases and regulators 10 ...
Page 13: ... Hardware Reference 13 ...
Page 25: ... Chemical Ionization Theory 3RVLWLYH WKHRU 25 ...
Page 31: ... Chemical Ionization Theory 1HJDWLYH WKHRU 31 ...
Page 34: ...34 ...
Page 92: ...92 ...
Page 114: ...114 ...
Page 120: ...120 ...
Page 122: ...H Manual Part Number G1999 90001 Copyright 1998 Hewlett Packard Printed in USA 2 98 ...