Troubleshooting
1RSHDNV
No PFDTD peaks in PCI
•
Incorrect reagent gas. There
are
no PCI PFDTD peaks created with isobutane
or ammonia. Switch to methane.
•
Analyzer not sealed (big air leak)
•
No calibrant in vial
•
Defective calibration valve
•
Air leak in carrier or reagent gas path
No reagent gas peaks in NCI
•
Reagent gases do not ionize in NCI; look for background ions instead.
•
Verify tune parameters
•
If no background ions are visible, go back to methane PCI
No PFDTD calibrant peaks in NCI
•
Look for background ions
•
Verify tune parameters
•
Go back to methane PCI
No sample peaks in NCI
•
Look for background ions
•
Go back to methane PCI
Large peak at
m/z
283 in NCI OFN spectrum
•
Look for background ions
•
Find and fix your small air leak