Troubleshooting
3RRUYDFXXPZLWKRXWUHDJHQWJDVIORZ
75
Poor vacuum without reagent gas flow
Possible Cause
Excess water in the background.
Action
Scan from 10 – 40 amu. A large peak at
m/z
19 (>
m/z
17) indicates water in the
background. If water is present, allow the instrument to bake out more and flow
reagent gas through the lines to purge any accumulated water.
Possible Cause
Air leak.
Action
Run Methane Pre-tune. See page 54. A visible peak at
m/z
32 indicates air in the
system. Check for and correct any leaks. See the
Leaks
section at the beginning of
this chapter.
Possible Cause
The foreline pump is not working properly.
Action
Replace the pump oil. If that does not help, it may be necessary to replace the pump.
Contact your local Hewlett-Packard Customer Engineer.
Possible Cause
The turbo pump is not working properly.
Action
Check the pump speed. It should be at least 95%. Contact your local Hewlett-
Packard service representative
& $ 8 7 , 2 1
Use of ammonia as reagent gas can shorten the life of the foreline pump oil and possibly
of the foreline pump itself. See the Maintenance chapter in this manual.
Summary of Contents for 5973 MSD
Page 1: ...H Hardware Manual HP 5973 MSD PCI NCI ...
Page 6: ...6 ...
Page 7: ...1 General Information Specifications 9 Reagent gases and regulators 10 ...
Page 13: ... Hardware Reference 13 ...
Page 25: ... Chemical Ionization Theory 3RVLWLYH WKHRU 25 ...
Page 31: ... Chemical Ionization Theory 1HJDWLYH WKHRU 31 ...
Page 34: ...34 ...
Page 92: ...92 ...
Page 114: ...114 ...
Page 120: ...120 ...
Page 122: ...H Manual Part Number G1999 90001 Copyright 1998 Hewlett Packard Printed in USA 2 98 ...