Program message terminator
The program instructions within a data message are executed after the
program message terminator is received. The terminator may be either an NL
(New Line) character, an EOI (End-Or-Identify) asserted in the HP-IB
interface, or a combination of the two. Asserting the EOI sets the EOI control
line low on the last byte of the data message. The NL character is an ASCII
linefeed (decimal 10).
New Line Terminator Functions
The NL (New Line) terminator has the same function as an EOS (End Of String)
and EOT (End Of Text) terminator.
Selecting multiple subsystems
You can send multiple program commands and program queries for different
subsystems on the same line by separating each command with a semicolon.
The colon following the semicolon enables you to enter a new subsystem. For
example:
<program mnemonic><data>;
:<program mnemonic><data><terminator>
:TRIGGER:MODE NORMAL;:TIMEBASE:RANGE 1
Combining Compound and Simple Commands
Multiple commands may be any combination of compound and simple
commands.
Introduction to Programming
Program message terminator
20
Summary of Contents for 54620A
Page 6: ...6 ...
Page 9: ...1 Introduction to Programming ...
Page 21: ...2 Programming Getting Started ...
Page 35: ...3 Programming over HP IB ...
Page 40: ...40 ...
Page 41: ...4 Programming over RS 232 C ...
Page 48: ...48 ...
Page 49: ...5 Programming and Documentation Conventions ...
Page 53: ...Programming and Documentation Conventions The command tree 53 ...
Page 60: ...60 ...
Page 61: ...6 Status Reporting ...
Page 63: ...Status Reporting Data Structures Figure 4 Status Reporting 63 ...
Page 68: ...68 ...
Page 69: ...7 Installing and Using the Programmer s Reference ...
Page 76: ...76 ...