Chapter 4 Command Reference
:MEASure Subsystem
Programming Guide
4-11
4
Table 4-1. Summary of the Measurement Instruction Commands
The <source_list> parameter has the same syntax as SCPI
<channel_list>. For example, the Frequency function uses (@1) to specify
channel 1.
If the instrument receives an unexpected parameter, it processes the
command, ignoring the unexpected parameter, and sets the “Command
Warning” bit of the Data Questionable status reporting structure.
The response format for :MEASure query, :READ?, and :FETCh? is ASCII
data. If no valid data is available, the Counter generates error -230 (Data
corrupt or stale).
Command
Description
:MEASure query
This command is the simplest to use, but it allows little flexibility.
This command lets the Counter configure
itself for an optimal
measurement, initiate the measurement, and return the result;
i.e., it provides a complete measurement sequence (:MEAS
query is equivalent to the :CONF, :INIT, :FETC? command
sequence, but with no flexibility.)
:CONFigure
:READ?
The combined use of these two commands allows for more
control when the Counter performs a measurement, initiates a
measurement, and returns the result. Use this command
sequence when you want to customize the configuration
between the measurement setup and acquisition.
:CONFigure
:INITiate
:FETCh?
This combination of commands provides the most flexibility the
Measure Instructions allow. This command sequence configures
the Counter, initiates the measurement as specified, and returns
the result.
Summary of Contents for 53150A Series
Page 1: ... Programming Guide HP 53150A 151A 152A Microwave Frequency Counter ...
Page 2: ......
Page 10: ...Contents x Programming Guide ...
Page 11: ...1 Before You Start ...
Page 20: ...Chapter 1 Before You Start Related Documentation 1 10 Programming Guide 1 ...
Page 21: ...2 Command Summary A Quick Reference ...
Page 41: ...3 Programming Your Counter for Remote Operation ...
Page 101: ...4 Command Reference A Dictionary ...
Page 162: ...Chapter 4 Command Reference Common Commands 4 62 Programming Guide 4 ...
Page 163: ...5 Errors ...
Page 174: ...Chapter 5 Errors Error Types 5 12 Programming Guide 5 ...
Page 184: ......
Page 186: ... Manual Part Number 53150 90002 Printed in U S A NOVEMBER 1997 ...