Chapter 2 Calibration Procedures
Adjustment Procedures
20 MHz
Flatness Adjustment
This procedure manually adjusts the
20
MHz flatness.
The
adjustment
is made with a 100 m V signal applied at
20
MHz. You
will
make the
flatness adjustment using
C304
which is located on the main circuit
board
(see
below).
The signal measured for this test is a dc voltage taken at the rear-panel
dc output jack. This voltage is normalized to
-1
Vdc and is proportional
to the
RMS
input voltage
(-1
Vdc corresponds to a full-scale meter
deflection).
HP3400B
Range
=:=! :::::::
20 MHz
Flatness
�stm8nt (C304)
::::::: -c:=
Calibrator
Calibrator
Oulpu! Voltage Output Frequency
DCOutpu!
(DMM Reeding)
0. 1
V
0. 1
V
20 MHz
-1 .0 125
V to
-0. 9875
V
t HP 3400B: Select the
0. 1
volt range.
2
Calibrator: Select a
0. 1
volt ac signal at
20
MHz (use the
"Wideband" output on the Fluke 5700A).
3
External Multimeter: Select the
10
Vdc range with 51.<.! digits of
resolution.
4
Remove only the right-side cover from the
HP
3400B (as viewed
from the front panel). It is important that you
do IWt
remove the
left-side. toP. or bottom covers. Adjust
C304
until the voltage
measured at the rear-panel dc output jack is within the limits
shown in the table above.
33
2
Summary of Contents for 3400B
Page 11: ... 1 1 Quick Start 1 ...
Page 19: ... 10 0 __ 2 Calibration Procedures 2 ...
Page 39: ... 3 Theory of Operation 3 ...
Page 48: ... 4 4 Service ...
Page 52: ... 10 _ _ 5 Replaceable Parts 5 ...
Page 64: ... 6 Backdating 6 ...
Page 66: ... 7 Schematics 7 ...
Page 69: ...C N z N a u r 0 0 e 5 is c 1 I __ __ _ _____1__ ...