Chapter
2
calibration Procedures
Adj ustment Proced ures
Tenth
ScaJe
Galn Adjustment (R3t6)
c=
-C>-
c:::::J -c::>-
-c::>-
:8=
�c:
HP3400B
Range
0.01 V
-<J-
calibrator
calibrator
Output Voltage Output Frequency
O.OOt
V
1
kHz
1 HP 3400B: Select the 0.01 volt range.
DC Output
(DMM Reacing)
-o.
1019V
to
-0.0981 V
2
Calibrator: Select a 0.001 volt ae signal at
I kHz
(use the
"Wideband" output on the Fluke 5700A).
3
External Multimeter: Select the 10 Vde range with 5\.2 digits of
resolution.
4
Remove only the top cover from the
HP
3400B (it is important
that you
do
Mt
remove the side or bottom covers). Adjust
Ral6
until the de voltage measured at the rear-panel de output jaek is
within the limits shown in the table above.
Note Since the full scale and tenth scale aqjustments are interdependent,
you
slwuld repeat both aqjustments again. Repeat the procedures again,
slorting with step 1 for the full scale aqjustment, on page
30.
31
2