Performa nce Verification Proced ures
1 HP
3400B:
Select the
1
volt range.
2
Calibrator: Select a
O. 1
volt ac signal at
10 Hz.
3
External Multimeter: Select the
10
Vdc range with
51-\!
digits of
resolution.
4
Verify that the de voltage measured at the rear-panel de output
jack is within the limits shown in the table below.
1\
Continue checking the de output as you adjust the calibrator's
output voltage and frequency to each of the remaining values
shown in the table below.
HP 3400B
calibrator
calibrator
DC Output
Range
Output Vo�
Output Frequency
(DMM Reading)
I V
0.1 V
1 0 Hz
-<l.1 5 V to -<l. 05 V
I V
0.1 V
so
Hz
-1.075 V to -<l.0925 V
I V
0.1 V
1 MHz
-1.075 V to -<l.0925 V
I V
0.1 V
2 MHz
-<l.1 2 V to -<l.08 V
I V
0.1 V
3 MHz
-<l.1 3 V to -<l.07 V
I V
0.1 V
10 MHz
-<l.1 5 V to -<l.05 V
I V
0.1 V
20 MHz
-<l.1 5 V to -<l.05 V
24
•
•
•
•
•
•
•
•
-,