Chapter 2
calibration
Procedures
Performa nce Verification Proced ures
Crest Factor Check
This test verifies the voltmeter's crest factor performance. The signal
measured for this test is a dc voltage taken at the rear-panel dc output
jack. This voltage is normalized to -Q.9487 Vdc and is proportional to
the
RMS
input voltage (-Q.9487 Vdc corresponds to a full·scale meter
deflection).
--11ms�
o v
--
HP
34006
Range
0.3V
Soorce
Configuration
Vou1a1.5
V.
Offse1=7 5O
mV.
Pulse
Ou1put
Wldth.1 mSec.
Frequoncy=10 Hz. Modo=Pu!se
1 HP 3400B:
Select the
0.3
volt range.
DC 0u1pu1
(DMM Roading)
-1.05
V
10 -0.95
V
2
Source: Program the source to output a pulse train with the
following characteristics: Vout=
1. 5
volta, offset=750 m V,
pulse output width=
1
mSec, frequency=
10 Hz,
mode=pulse.
3
External Multimeter: Select the
10
Vdc range with
5\.2
digits of
resolution.
4
Verify that the dc voltage measured at the rear·panel dc output
jack is within the limits shown in the table above.
27
2