TB 9-6625-2049-35
19
Table 5. Dc Offset - Continued
Test instrument
DC OFFSET
Multimeter indications
(V dc)
settings
Min
Max
14.99 mV
0.01493
0.01505
-14.99 mV
-0.01505
-0.01493
4.999 mV
0.004979
0.005019
-4.999 mV
-0.005019
-0.004979
1.499 mV
0.001479
0.001519
-1.499 mV
-0.001519
-0.001479
NOTE
Perform (8) through (11) below for option 002 only.
(8) Remove 50
Ω
feedthrough termination from equipment setup.
(9) Press
SIGNAL
key on.
(10) Enter
20 VOLT
using
DATA
keys. Multimeter will indicate between 19.775
and 20.225 V dc.
(11) Enter
-20 VOLT
using
DATA
keys. Multimeter will indicate between -19.775
and -20.225 V dc.
b. Adjustments
. No adjustments can be made.
18. Dc Offset with Ac
a. Performance Check
(1) Connect
SIGNAL
to multimeter
INPUT HI
and
LO
using 50
Ω
feedthrough
termination.
(2) Press keys and enter values using
DATA
keys as listed in (a) through (i) below:
(a)
SIGNAL
off (option 002 not lit).
(b)
FUNCTION
sine wave.
(c)
ENTRY FREQ
.
(d)
20.999 999 999
MHz
.
(e)
ENTRY AMPTD
.
(f)
1
VOLT
.
(g)
ENTRY
DC
OFFSET
.
(h)
4.5
VOLT
.
(i)
AMPTD CAL
.
(3) Multimeter will indicate between 4.35 and 4.65 V dc.