TB 9-6625-2049-35
20
(4) Enter -
4.5 VOLT
using
DATA
keys. Multimeter will indicate between -4.35 and
-4.65 V dc.
(5) Press
ENTRY FREQ
key and enter
999.9 kHz
using
DATA
keys. Multimeter
will indicate between -4.44 and -4.56 V dc.
(6) Press
ENTRY
DC
OFFSET
key and enter
4.5 VOLT
using
DATA
keys.
Multimeter will indicate between 4.44 and 4.56 V dc.
(7) Press
FUNCTION
square wave key. Multimeter will indicate between 4.44 and
4.56 V dc.
(8) Enter -
4.5 VOLT
using
DATA
keys. Multimeter will indicate between -4.44 and
-4.56 V dc.
(9) Press
ENTRY FREQ
key and enter
9.9999 MHz
using
DATA
keys.
Multimeter will indicate between -4.35 and -4.65 V dc.
(10) Press
FUNCTION
triangle key and enter
9.9
kHz
using
DATA
keys.
(11) Multimeter will indicate between -4.44 and -4.56 V dc.
(12) Press
FUNCTION
positive ramp key. Multimeter will indicate between -4.38
and -4.62 V dc.
b. Adjustments
. No adjustments can be made.
19. High Voltage Output Option 002 only
a. Performance Check
(1) Connect
SIGNAL
to multimeter
INPUT HI
and
LO
.
(2) Press keys and enter values using
DATA
keys as listed in (a) through (i) below:
(a)
FUNCTION
sine wave.
(b)
ENTRY FREQ
.
(c)
2
kHz
.
(d)
SIGNAL
on.
(e)
ENTRY DC OFFSET
.
(f)
0
VOLT
.
(g)
ENTRY AMPTD
.
(h)
14.14
V RMS
.
(i)
AMPTD CAL
.
(3) Multimeter will indicate between 13.86 and 14.42 V ac.
(4) Press keys and enter values using
DATA
keys as listed in (a) through (e) below: