TB 9-6625-2049-35
22
Figure 5. Power supply assembly A2 – test instrument top view.
(5)
Press
presently active
FUNCTION
key to remove ac output and activate
ENTRY DC OFFSET
key (lit).
(6) Enter
0 VOLT
using
DATA
keys and press
AMPTD CAL
key.
(7) Connect multimeter
INPUT HI
to TP AMP OUT (fig. 3) and
LO
to circuit board
ground.
NOTE
The voltages measured in (1) through (3) above may be
adjusted out of tolerance by (8) through (11) below. This is not
a cause for concern.
(8) Adjust R40 (fig. 3) until digital multimeter indicates less than 5 mV dc (R).
(9) Connect multimeter
INPUT HI
and
LO
to TI
SIGNAL
. Do not use 50
Ω
feedthrough termination.
(10) Enter
5
VOLT
using
DATA
keys.
(11)
Press
AMPTD CAL
key. Adjust R22 (fig. 5) until multimeter indicates 10.000
V dc.
(12) Enter
-5
VOLT
using
DATA
keys.
(13) Multimeter will indicate between -9.985 and -10.015 V dc.
21. Final Procedure
a
. Deenergize and disconnect all equipment.
b
. Annotate and affix DA label/form in accordance with TB 750-25.