HOVER
MATT
®
USER MANUAL
9
Rev A
HMManual
Product Specifications/Required Accessories
• Classification:
EQUIPMENT NOT SUITABLE FOR USE IN THE PRESENCE OF A FLAMMABLE
ANESTHETIC MIXTURE WITH AIR, OR WITH OXYGEN OR NITROUS OXIDE.
Not for use with Oxygen or Oxygen Enriched Atmospheres.
Type of Protection against electric shock: CLASS I EQUIPMENT
Degree of protection against electric shock: TYPE BF APPLIED PART
Protection against ingress of water: Ordinary (not protected).
Mode of operation: CONTINUOUS OPERATION
To remove supply mains, unplug equipment from wall
• Patient Weight Limit:
1,200lbs / 544kg
• Use Temperature:
50° to 104° F (10° to 40° C)
• Use Humidity:
10% to 70% Non-Condensing
• Storage/Shipping Temperature: -40° to 176° F (-40° to 80° C)
• Storage/Shipping Humidity:
10% to 70% Non-Condensing
• Power Input:
120 V~, 60 Hz, 10 A (North American version)
• Air Supply Dimensions:
12.5 x 7 x 7 inches (31.75 x 17.8 x 17.8 cm)
• Air Supply Weight:
11 lbs. (5 kg)
• Air Supply Material:
Fire Retardant ABS/Stainless Steel
230 V~, 50 Hz, 6 A (Australian Version)
• Power Cord Length:
10 metre