Electromagnetic Compatibility Chart
HOVER
MATT
®
USER MANUAL
12
Rev A
HMManual
For HTAIR-2300 ONLY
EMC: Table 2
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The HoverTech International Air Supply is intended for use in the electromagnetic environment
specified below. The customer or the user of the HoverTech International Air Supply should ensure
that it is used in such an environment.
Immunity Test
IEC 60601 Test
Level
Compliance Level
Electromagnetic Environment-
Guidance
Electrostatic
Discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%.
Electrical fast
Transient/burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for input-
output lines
± 2 kV for supply
mains
± 1 kV for
input/output lines
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
< 5%
U
T
(> 95% dip in
U
T
)
For 0,5 cycle
40%
U
T
(60% dip in
U
T
)
For 5 cycles
70%
U
T
(30% dip in
U
T
)
For 25 cycles
< 5%
U
T
(>95% dip in
U
T
) for
5 seconds
< 5%
U
T
(> 95% dip in
U
T
)
For 0,5 cycle
40%
U
T
(60% dip in
U
T
)
For 5 cycles
70%
U
T
(30% dip in
U
T
)
For 25 cycles
< 5%
U
T
(>95% dip in
U
T
) for
5 seconds
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the HoverTech International Air
Supply requires continued
operation during mains
interruptions, it is recommended
that the HoverTech International
Air Supply be powered from an
uninterruptible power supply or a
battery.
Power Frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic
of a typical commercial or
hospital environment.
NOTE
:
U
T
is the AC mains voltage prior to application of the test level