SECTION 6 TROUBLESHOOTING
430-00587-008 (Rev. 2/05)
6 - 12
Plum XL Series
When the infusion system is in service mode and the control knob is set to the SET VTBI
position, the software revision number can be viewed. The decimal point does not appear
in the software revision number display, but is implied after the first digit. For example,
if the display shows 105, the software revision number is 1.05. The software revision
number may be necessary when contacting Hospira.
When the infusion system is in service mode and the control knob is set to the RUN
position, the run time and battery run time can be viewed. In the run time and battery
run time display, large digits indicate the total infusion system run time in tens of hours
and the small digits indicate the battery run time in tens of hours. For example: if the
large digits indicate 245 and the small digits indicate 79, the infusion system has been
operated for a total of 2,450 hours and has also been operated on battery for 790 of those
2,450 hours.
When replacing the battery, it may be desirable to clear the battery run time. To clear the
battery run time, simultaneously press and hold the [PRI/SEC] key and the [BACKPRIME]
key for four seconds. The small digits flash and an audible tone sounds four times at
a once-per-second rate. After four seconds, the battery run time is cleared. The total
infusion system time cannot be cleared.
PARAMETER PROGRAMMING (XLM WITH DATAPORT)
When the infusion system is in service mode and the control knob is set to HOLD/RESET,
three sub-modes can be viewed and changed. Each of these three sub-modes are used
to change the value of an operational parameter of the infusion system as shown in
Table 6-4, Sub-Modes of Parameter Programming
The first sub-mode (communication selection) is the default sub-mode of parameter
programming. Subsequently pressing the [BACKPRIME] key will change it to the second
sub-mode (soft ID), the third sub-mode (channel label), then to the first sub-mode. Within
a sub-mode, the small digits indicate its index while the large digits indicate the current
value of the parameter to be viewed and programmed.
Table 6-4.
Sub-Modes of Parameter Programming
Sub-Mode Name
Purpose
Sub-Mode
Index
Communication Selection
To select the communication with
the host computer
1
Soft ID
To enter the soft ID of the infusion system
2
Channel Label
To enter the channel label of the infusion system
3