18
2. LED Lights and Audible Alarm Safeties
The "POWER OK" LED indicates proper control voltage and will remain on unless a
control voltage problem occurs. For further details, see "II.B. Sequence of Operation."
Sequence Step
LED
Energized
Components
Min.
Max.
Fill Cycle
WTRIN
WV
-
-
Ice Purge Cycle
GM
GM, EH
5 min.
5 min.
Freeze Cycle (with refill)
GM, COMP,
WTRIN* (refill),
GM, EH, PDS, Comp,
FMR,
WV* (refill)
-
*On until UF/S closes.
Alarm sounds after
90 sec.
1-in-1 Drain Cycle
(S1 dip switch 4 off)
FLUSH (Drain)
GM, COMP
(only if IS not
satisfied)
DV
Comp, GM, EH, PDS,
FMR (only if IS not
satisfied)
2 sec.
2 sec.
1-in-12 Drain Cycle
(S1 dip switch 4 on)
FLUSH (Drain) DV
10 min.
10 min.
The built-in alarm safeties shut down the unit.
No. of Beeps
(every 5 sec.)
Type of Alarm
Reset Options
1
Low Water Safety
UF/S open > 90 seconds after WV energized.
Automatic reset once water supply is restored
and UF/S closes.
2
Control Switch
In "DRAIN" position longer than 15 minutes.
Automatic reset once the control switch is
moved to the "ICE" position.
3
High-Pressure Switch: N/A on this model.
See "III.B.2. SRC-10H" for high-pressure
switch operation.
4
High-Pressure Switch: N/A on this model.
5
Freeze Timer
WV off > 30 minutes since last WV activation.
Manual reset. Turn power off and on again.
6
Low Voltage
(92VAC±5% or less)
"POWER OK" LED turns off if voltage
protection operates.
The control voltage safeties automatically
reset when voltage is corrected.
7
High Voltage
(147VAC±5% or more)
8
Gear Motor
GMPR contacts fail to close.
Manual reset. Turn power off and on again.
9
Infrared Sensor (S1 dip switch 7)
MBC actuator paddle engaged.
Manual reset. Turn power off and on again.
Legend:
DV
–drain valve;
EH
–evaporator heater;
FMR
–fan motor-remote;
GM
–gear
motor;
IS
–infrared sensor bin control;
MBC
–mechanical backup bin control;
PDS
-pump-down solenoid valve;
WV
–inlet water valve