5.2

Temperature display check

1.

Select Temp Set with MODE key on the Simulator.

2.

Start outputting simulation input value with OUTPUT
key.

3.

Input temperature simulation value in the order of
0.0°C, 25.0°C, 30.0°C, 60.0°C and 100.0°C to the
Simulator, then check the value on the instrument.

6

Specifications

Note

This Simulator supplies simulation input specific to HORIBA
conductivity meter. Therefore, the simulator output cannot be
measured by general-purpose digital multi-meter.

Output

Output 
parameter

Output value

 Accuracy (Varies depending on 
temperature used.)

±5°C of the 
temperature in X-
52 delivery 
inspection(*)

5°C to 40°C

COND 
Span

19.00 S/m 

±0.1 S/m 

±0.8 S/m ±1digit

1.900 S/m to 
1.900 mS/m

±0.1% ± 1digit 

±0.2% ±1digit

COND 
Linear

10.00 S/m 

±0.1 S/m 

±0.4 S/m ±1digit

1.000 S/m to 
10.00 

µ

S/m

±0.1% ± 1digit 

±0.2% ± 1digit

0.000 mS/m

-

-

Tempera-
ture

0.0/ 25.0/ 
30.0/ 60.0/ 
100.0

°

C

±0.1

°

C ±0.2

°

C

Ambient 
temp.

5°C to 40°C

Ambient 
humidity

0% to 80% (No dew condensation)

Power

Two AA alkaline dry-cell batteries (Approx. 20 hours with 
continuous operation)

Dimensi
ons

170(H) 

×

 89(W) 

×

 40(D) mm

(*):

Temperature in X-52 delivery inspection is described
in the inspection report

.

1. Select Temp Set.

2. Output simulation 

3. Toggle simulation 

input value.

input value.