3 Basic Operation
28
Atmospheric pressure compensation (DO)
Differences in the atmospheric pressure of the measurement location influence the Dissolved
Oxygen (DO) measurement. By setting (input) the actual atmospheric pressure of the
measurement location into the control unit, it is possible to standardize the measured
Dissolved Oxygen (DO) value to a value at the standard atmospheric pressure (1013 hPa).
Reference
“ 6.4.4 Atmospheric pressure compensation (DO) - Relation of altitude and air pressure ” (page 107)
Note
If air bubble in the internal solution is large, change the internal solution.
Calibrate at the place of same atmospheric pressure before the measurement.
1.
Press the control unit’s POWER key for about 1 second to turn the power ON.
The "MEASUREMENT" screen appears after about 10 seconds.
Note
The operation keys are designed to operate using the pad of a finger, sharp objects can tear the
control unit cover damaging the operation keys.
2.
Press the right (
) key to switch the display to the "SETTINGS" screen.
3.
Press the down (
) key to move the cursor to "Compensation", then press the
ENTER key.
4.
Press the down (
) key to move the cursor to "Cond Temp", then press the ENTER
key to toggle the setting between "OFF" and "Input mode".
Default: OFF