Dissolved oxygen measurement
40
• Setting salinity correction (default: 0.0 PPT)
Because dissolved oxygen is affected by the salinity concentration, you need to correct it
at measurement. Set the salinity concentration of the sample beforehand.
You can select arbitrary numeric value or sea concentration (35.0 PPT). If the measured
target does not contain saline matter, set to 0.0 PPT.
1. Press the
key to change the setting mode.
2. Press the
keys to select "DO" (DO correction setting) and then press the
key.
3. Press the
keys to select "SAL" (salinity correction setting) and then
press the
key.
4. Enter the salinity coefficient and press the
key.
The setting range is 0.0 PPT to 40.0 PPT.
To return to the measurement mode, press the
key.
key
key
key
Enter a
number
key