When an error message appears
88
●
ERR No.0006
Maximum calibration points exceeded
Attempted to perform 6th point calibration during pH calibration.
●
ERR No.0007
Cannot identify standard solution
Unable to automatically detect the standard solution during pH calibration.
●
ERR No.0008
Calibration interval error
・
More than the set number of days has elapsed since calibration was last performed.
・
The calibration interval alarm is "ON" and calibration has not been performed.
Cause
How to solve problem
6th point calibration is
attempted.
Up to five points can be calibrated.
Cause
How to solve problem
There is a problem with the
standard solution.
Use new standard solution.
Settings about the standard
solution do not match the
instrument.
Check if the instrument settings and the specifications of
the standard solution are compatible.
The electrode is dirty.
Clean the electrode.
The electrode is cracked.
Replace the electrode.
The standard solution
concentration is fluctuating.
Replace the internal solution in the reference electrode.
Electrode is not immersed
enough to cover liquid junction.
The electrode must be immersed up to the liquid junction.
As a guide, immerse to at least 3 cm from the tip of the
electrode.
Cause
How to solve problem
Calibration has not been
performed for the set number of
days of the calibration interval or
longer.
Perform calibration.
The calibration interval alarm is
"ON" and calibration has not
been performed.
Perform calibration.
Summary of Contents for LAQUAact-PD110
Page 1: ...Handheld pH ORP Dissolved Oxygen Meter LAQUAact PD110 Instruction Manual CODE GZ0000427800...
Page 2: ......
Page 12: ...X M E M O...
Page 100: ...M E M O 84...
Page 110: ...M E M O 94...
Page 117: ...2 Miyanohigashi Kisshoin Minami ku Kyoto 601 8510 Japan http www horiba com...